Reconditioning and Disinfection ................20 Technical Safety Inspection ..................21 Lifespan ......................... 21 Electric-Magnetic Compatibility ................21 Entsorgung von Verpackung und Elektro-Altgeräten ........22 Technical Data ....................23 Distribution / Trading ....................23 User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Patients with impaired sensibility in their hands or feet (e.g. patients with polyneuropathies) Warning: The treatment system may only be powered by the HIDREX wall adapter (to recognize at HIDREX nameplate with Serial-No.). This power supply is especially designed for this device.
HIDREX Company. Never open the device; there are no control elements inside. The system may only be serviced by the HIDREX Company. Wash the supplied towels on a regular basis together with your ...
Intended Use / Mechanism of Action During HIDREX treatment, a current flows through the body regions un- der treatment. The water bath mediates this current flow. The skin areas inside the treatment water will thereby secrete less sweat. Although treatment success has been validated in numerous medical studies, there is still no completely satisfactory scientific explanation for the mechanism of action.
10 treatments. Phase 2: Long term treatment (maintenance therapy) is indicated because the HIDREX treatment effect is reversible. Patients should conduct maintenance therapy sessions by themselves at home and with their own unit. Depending on the severity of the condition, maintenance therapy involves one to three weekly sessions of approximately 15 minutes each.
(AX electrodes) made of foam and come equipped with hook and loop tapes (used with the AX elec- trodes) includes 2 additional electrodes each, towels, and a dedicated cable set User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Button: Time - lower dose Setup: confirm primary settings increase dose decrease time increase time Setup: Memory function indication memory function active Immersion monitoring (Dive) User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
(the backs of the hands and backs of the feet should not be covered with wa- ter!). User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Please ascertain that the sponge material is completely soaked (do not squeeze). Now secure the applicator with the hook and loop fasteners. Important: The treatment with face mask is possible only in the MEMO mode! User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Primary Parameter Setup (Setup Mode) Depending on the specific HIDREX system you own, you can choose be- tween various options before you initiate a treatment. How you select and store these parameters with the GP450 is described in the following paragraphs.
Page 12
Adjust the system in accordance with your doctor’s instructions. User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Select the dose in accordance with your doctor’s orders or use the table of factory- preset treatment parameters for guidance (see page 14 in this manual). User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Page 14
In the factory, your HIDREX system is preset as follows: Option Parameter MEMO DOSIS: 0 Volt TIME: 15 minutes In order to avoid skin irritation by excessive dose settings, we recom- mend that you assess your individual direct current thresholds before you start the first treatment session.
Page 15
Please keep your hands and feet in the water bath (or the axillary applicators in place, respectively) until the dose display reads “End“. Switch the device off. User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Important Advise for Conducting Treatments For your safety, your HIDREX treatment system comes equipped with several protective circuits. Immersion monitoring Until the treatment current path is closed (immersion monitoring) by im- mersion of the palms, the dose cannot be increased to any value above zero.
Important: Make absolutely sure to pay attention to the list of contraindica- tions! Excessive dose settings can lead to burns! General Information This chapter contains important instructions about your HIDREX system. Special Remark Our responsibility for system safety, functionality, and reliability applies only if any maintenance and servicing is exclusively performed by ourselves or by personnel authorized by us.
Dr. Trippen (desomed): Biguamed B. Braun: Melesept SF (0.5%/5%), Hexaquart S (1.5%/5%), Meliseptol (undiluted) Operational Check Proceed with the following steps for an operational check of your HIDREX system: 1) Set up the therapy system as you would for a treatment.
In that case, try non-carbonated mineral or table water instead. Device leakage currents comply with standards -- the system provides protection against electrical shock (Type B); device is insulated (floating) (Type F). User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Please do not ship towels, sponge cushions, or the facial mask. Please send all electrical accessories (safety wall adapter, treatment and AX electrodes, and dual connecting cable set) together with the HIDREX control unit. Applicable Regulations and Legal Requirements...
Each successful reconditioning by the manufacturer extends the lifespan of the medical device by 2 years. If the HIDREX iontophoresis sys- tem is reconditioned for the same patient, the treatment trays or carrying case (depending on their condition) do not necessarily have to be replaced.
You help to avoid with it that incriminating materials reach into the environment (ElektroG). WEEE-Reg.-Nr.: DE 42510094 Hersteller-Nr. Duales System Interseroh: 134502 (VerpackV) User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Output current max. 0,5 A Max. output power 6 VA Distribution / Trading United Kingdom: Avanor Healthcare Ltd Unit 19 Mahoney Green, Rackheath, Norwich, NR13 6JY, UK Phone: 0845 643 9545 www.avanor.co.uk User Manual Tap Water Iontophoresis System GS400 Rev. 1.0 / 2011-10-10...
Need help?
Do you have a question about the GS400 and is the answer not in the manual?
Questions and answers