Checking for Surface Contam ination 5 Maintenance Calibration Troubleshooting Service 6 Basics of Radiation and Its Measurem ent Ionizing Radiation Radiation Measurem ent Units Appendix A Technical Specifications Appendix B Technical Update Warranty Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
The num ber of counts detected by the palm RAD 907 varies from reading to reading due to the random nature of radioactivity. A reading is ex pressed m ore accurately as an Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
If you ex pect to not use the palm RAD 907 for longer than one m onth, rem ove the battery to avoid dam age from battery corrosion. Change the battery prom ptly when the battery indicator appears on the display. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
(LCD) (4). You control which unit of m easurem ent is shown by using the m ode switch. Whenever the palm RAD 907 is operating, the red count light (1) flashes each tim e a count (an ionizing event) is detected. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
The current unit of m easurem ent (I)- - CPM, CPS, m R/ hr or Sv/ hr- - is displayed to the right of the num eric display. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
Utility Menu. For m ore inform ation, see "Taking a Tim ed Count," Using the Alert, and The Utility Menu in Chapter 3 and "Calibration" in Chapter 5. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
If the probe is not connected when you turn the palm RAD 907 EXP on, the instrum ent will not function properly. CAUTION: Do not rem ove the probe while the instrum ent is on. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
Interfacing to an Ex ternal Device in Chapter 3. The probe port on the end panel of the palm RAD 907 EXP allows you to plug the ex ternal probe into the instrum ent. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
The m ost im m ediate indicators of the radiation level are the count light, the audio beep, and the alert. It takes three seconds before an increase is shown on the num eric display in the dose rate m odes. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
R/ hr; when the tim ed period is over, the total is displayed whenever you switch back to Total/ Tim er. The hourglass indicator is shown on the display in any m ode setting; it is blinking while the tim er is totaling counts. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
30 seconds. In order to give a quicker response to changes, when the radiation level ex ceeds 6,000 CPM in any 30- second period, the reading is based on the previous 6 seconds, and Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
The best alert level is one that rarely gives a false alarm , yet warns you when the radiation is higher than norm al. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
Geiger tube detects a count. A cable with an RS- 232 connector for an IBM PC- com patible com puter serial port and accom panying software are available from Berkeley Nucleonics Corporation. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
If you suspect an increase that is too sm all to detect with a short tim ed reading, you can take a longer count (for ex am ple 6, 12, or 24 hours). Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
CAUTION: Never touch the palm RAD 907 to a surface that m ay be contam inated. You m ay contam inate the instrum ent. The rubber strips on the back can be replaced if they becom e contam inated. Replacem ent strips are supplied with the palm RAD 907. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
20% higher than the factory reading, the new calibration factor would be 120. The current calibration factor is displayed during the system check when the palm RAD 907 is first turned on. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
Page 20
Calibrat ing Elect ronically You can calibrate electronically using a pulse or function generator. Electronic calibration requires a cable with a 2.5 m m plug, with the tip carrying the signal. Follow these steps: Set the signal height to 3.3 volts (positive pulse) and the pulse width to 80 m icroseconds.
, it requires factory service continuous discharge replace the Geiger tube electrom agnetic field m ove the instrum ent away from possible sources of electrom agnetic or radio frequency radiation Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
Do not attem pt to repair the palm RAD 907; it contains no user- serviceable parts and you could void your warranty. CAUTION: Do not send a contam inated instrum ent for repair or calibration under any circum stances. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
6 Basics of Radiat ion and It s Measurem ent This chapter briefly tells what radiation is and how it is m easured. This inform ation is provided for users who are not already fam iliar with the subject. It is helpful in understanding how the palm RAD 907 works and in interpreting your readings.
. Microcuries (m illionths of a curie) and picocuries (trillionths of a curie) are also often used as units of m easurem ent. A bequerel (Bq) is equivalent to one disintegration per second. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...
Appendix A Technical Specificat ions Detector: Halogen- quenched Geiger- Mueller tube. Effective diam eter 1.75" (45 m m ). Mica window density 1.5- 2.0 m g/ cm EXP only: Sam e detector in anodized alum inum housing with black vinyl grip. 500 volt power supply is located in the probe head. Am phenol Tucal connectors.
Alert setting in any m ode to the desired setting. If you would like to have the corrected revision of firm ware installed, please return the unit to a factory authorized service center for reprogram m ing. Shanghai Mengliang Instrument Co., Ltd. TEL:+86-21-36308517 URL:www.menglianggroup.com...