Instructions for each of these modes and convenience features are provided in this manual. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Use of bichromatic differential absorbance values corrects for optical imperfections in the plastic wells, differences in well meniscus, turbidity, and lamp variation. A photo detector converts transmitted light energy into electrical signals, which are amplified and interpreted. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
The instrument should be surrounded by the following clearances: 8 cm around perimeter of unit and 8 cm on top. Do Not Operate Without Covers Do not operate this instrument with covers and panels removed. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
To avoid electric shock, the power cord protection ground conductor must be connected to ground. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
The serial connector of the reader is located on the right side of the rear panel. Interface: “DB- 9P” is configured for data terminal equipment (“D” connector with 9 male pin contacts.) Pin 1 is ground, Pin 2 is the transmit data output. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Check that the carrier moves in and back out of the instrument. Check that all 8 or 12 of the readings are 0.000 ± 0.005A. Poor repeatability of the absorbance readings for air indicates possible electronic problems. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Printer paper well (with paper roll cover) Cover Optical system (under cover) Printer paper slot Alpha-numeric display Keyboard Carrier track (with carrier) Place carrier groove over bar Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Pressing the CNTRL key causes the reader to ask for the locations and acceptance criteria of high and low controls to be used for automatic labeling. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Page 15
5 is 545nm, key 6 is 340nm 6 filter VIS: key 5 is 600nm, key 6 is 545nm The 1 and 0 keys are also used to answer YES and NO questions. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
98/79/EC IVD directive Physical Enclosure ...........ABS Plastic cover with metal base Dimensions ..........Approx. 9”x 12” x 3”, (23 x 30 x 8 cm) 13 lbs (5.9 Kg) Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Page 17
Optional Accessories Available from your dealer ......DRI-DYE Check Strips for instrument QC, replacement lamps, fuses, thermal paper, serial cable Design and instrument specifications are subject to change without notice. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Page 18
Notes: Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
11. Will a cutoff value be used to indicate negative samples or define an equivocal zone? If so, results less than what cutoff should be labeled as negative? 12. Will you be running partial strips? If so, you may want to assign an END location. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
You may change between reading 8 well strips and 12 well strips via the ALT key whenever the display indicates “Ready”. If a test is stored in the test menu, the strip type will also be saved. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
It also checks the individual control wells or the means for QC purposes. The user may select Correction factor and equivocal range options. These adjust the cutoff value to determine cutoff points for interpretation of subsequent specimens. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
CLEAR key twice to cancel the mode, and then begin again with the primary wavelength. Your choices will be printed, and the operating mode will resume. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
For calculating the standard curve, n=3, n6. If one well of a pair is deleted during an editing cycle, then the other well will represent the mean point and n=3, n2 and n5. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
In Point to Point and Multipoint % Abs modes, you may also choose to flag samples having absorbance values higher than the highest calibrator or lower than the lowest calibrator. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
“H”, “L”, or no flag are used to indicate high, low, or in range respectively. If your test is stored, the locations and acceptance criteria of the controls entered will also be stored for later recall. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Over bar in determine instrument carrier position. Ensure that these holes are not blocked or obscured by 12-Way Carrier 8-Way Carrier markings on (STANDARD) (OPTIONAL) the end of strip tabs. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Load the first strip into the right-most carrier position and press ENTER. While the strip is being automatically positioned for reading, the display will indicate “Positioning Strp”. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
If X*mNC is less than the Min C/O, the Min C/O will be used as the cutoff. The display will prompt “Min C/O =”. Enter the minimum cutoff value. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Page 29
Error messages will also print in the Interpretation column: H - indicates that the blank or the individual control exceeded the limit entered. L - indicates that the individual control was lower than the limit entered. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
The display shows: “Equiv Range Y/N”. If you answer YES, you will be able to enter an equivocal range based on factors of the cutoff value. The instrument will prompt “NEG<X*COV;X=”. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Page 31
Enter” is displayed, where X is either 1 or A, depending on the strip type. Load the first strip into the right-most carrier position and press ENTER. While the strip is being automatically positioned for reading, the display will indicate “Positioning Strp”. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
“# of Cals = ”. The number of calibrators must be less than or equal to eight (seven with a blank). The display then prompts for unit selection. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Page 33
The instrument will be ready for the next strip when it displays “Set Carrier”, “Then Press Enter”. To continue reading in the Regression mode, load the next strip and press the ENTER key. Press CLEAR twice to cancel the mode. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Note that the “Off Curve” flagging feature requires that the calibrators be entered and read from lowest concentration for calibrator #1 to highest concentration for the final calibrator. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
A calibrator absorbance. Except for the differences mentioned above, please follow the procedures for running the Point- to-Point mode to proceed in this mode. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Each time you store a test, the instrument will assign the next available test number. Remember to update your log. The instrument will record the date and time that each test was stored or last modified. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
If using a stored curve is inappropriate, you should re-run your calibrators upon test recall by selecting NO when “STORED CURVE Y/N” is prompted. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
“>10**6” is printed as the concentration value whenever the calculated result is greater than 6 digits and can not be properly printed in the concentration field. “CONTROLS FULL” is displayed if a fourth control location is attempted. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
When no differential filter is selected, the message “OFFSET ABS Y/N” is displayed. If you answer YES, you see “OFFSET ABS =”. Enter the value that should be subtracted from the observed blank absorbance. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
A monthly verification of instrument calibration and linearity is recommended. Such monitoring can be done using DRI-DYE Check Strips ® . There is more information about DRI-DYE Check Strips ® in the Appendix. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Use clean strips and follow the instructions for blanking and standardizing very closely. (Use of the appropriate blanking material is also very important...water alone is not always specified.) Do not read wells with bubbles or condensation. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Be careful not to soak liquids up under the keyboard overlay. If this happens, simply allow the keyboard to dry before resuming operation. Such a spill does not present a hazard, but may cause temporary malfunction of the keys. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Be certain that reagents are dispensed into the correct wells. Refer to the operating instructions for each mode regarding correct placement of the blank and calibrator materials. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Page 44
If only one setting is wrong, the printer will either print nonsense characters or nothing at all. Support is available to help you solve any technical problem or question with this instrument. Please call, fax, or write your dealer for immediate attention. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
2. Other engineering data supplied by Awareness Technology, Inc., Palm City, FL (1987-2006). 3. Data on Dri-Dye Check Strips provided by Awareness Technology, Inc., Palm City, FL (1988-2006). ® Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Page 46
Notes: Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
Appendix 4 provides contact information in the event the user has questions regarding instrument operation. Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
Check Strips can be ordered from your dealer. Available for the following wavelengths: DRI-DYE Check Strips-405 - for 405nm ® DRI-DYE ® Check Strips-450 - for 450nm DRI-DYE Check Strips-492 - for 492nm ® Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.
4.2 Appendix 2: Test Log Microstrip Reader (SN:________) User Test Menu Log Page ___ Test # Analyte: Stored By: Date: Notes: Awareness Technology, Inc. Stat Fax 303 Plus Operator’s Manual Rev. J...
• a description of the problem with as much detail as possible • printouts, which can be submitted to us by mail, fax or email. Stat Fax 303 Plus Operator’s Manual Rev. J Awareness Technology, Inc.