To keep your scale in top working order, the following
preventative maintenance measures apply:
1. This scale contains no moving parts in the
weighing portion, so recalibration should
never be required.
mechanical scales that require periodic
calibration to compensate for mechanical
wear.
However, we suggest that the
calibration should be checked annually or as
required.
2. Inspect cradle for cracks or loose mounting
hardware. Replace/repair as necessary. Do
not clean cradle with alcohol or other
drying agents.
3. Check scale's enclosure for damage or loose
or missing hardware.
necessary.
4. Inspect line cord for abrasion or other signs
of wear.
5. Do not expose scale to excessive water or
moisture.
6. Do not store the scale where heavy objects
may be set on it.
7. Replace batteries annually or as required to
insure optimum performance.
3.110

FACTORY RETURNS

If returning any components to the factory, please use
the original shipping containers designed for this
purpose.
You must call our service department
(630-653-3377 or 800-879-2001) for an RMA
(Returned Material Authorization) number, shipping
instructions and special address.
DO NOT SHIP ANYTHING TO THE
WHITE
PLAINS,
ADDRESS.
This will only cause delays and added expense to
you for the items to be reshipped to the factory in
Illinois.
RMA NUMBERS MUST BE OBTAINED FROM THE
FACTORY PRIOR TO RETURNING ANY ITEM.
This is different from
Replace/repair as
NEW
YORK,
11