    English Thank you for purchasing iStar. Please read this Operation Manual to ensure proper operation and maintenance of the product. Keep this Operation Manual available for future reference. - Contents - User and Intended Use ......................1 Cautions for handling and operation ..................2 1.
  • Page 4: Cautions For Handling And Operation

    Cautions for handling and operation Please read these instructions carefully and use only for intended purpose. Safety instructions are intended to avoid potential hazards that could result in personal injury or damage to the device. Safety instructions are classified as follows in accordance with the degree of risk. Classification Degree of Risk Hazard that could result in death or serious injury if the safety instructions...
  • Page 5 (EMC) (An emission increase or immunity decrease may occur). • iStar should not be used adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, iStar should be observed to verify normal operation in the configuration in which it will be used.
  • Page 6 • The maximum surface temperature of the handpiece might reach 45°C depending on use. • Extended continuous use may cause overheating of the handpiece motor which may result in an accident. Do not exceed 3 minutes of continual use and leave the handpiece off for at least 5 minutes before using again.
  • Page 7: Package Contents

    Package Contents Set components iStar Product name Quantity Handpiece motor Stand AC adapter Handpiece sleeve (100pcs.)
  • Page 8: Component Names

    (Alternate (1) POWER key operation) Alternatively, pressing and holding down the ON/ iStar OFF key rotates the handpiece motor while the button is being pressed. Releasing the ON/OFF key Operation Panel stops rotation. (Temporary rotating operation)
  • Page 9: Setup

    • Only use a disposable prophy angle. NOTICE • Any commercially available disposable prophy angle compatible to ANSI/ADA N. 85- 1:2004 can be used with the iStar. • Always wear safety glasses and dust mask during use. (1) Connecting 1) Turn off the handpiece motor.
  • Page 10: Preoperational Checks

    • Remove the handpiece sleeve from the handpiece motor before placing it on the battery charger. • The handpiece battery charging stand is exclusively for the iStar handpiece motor. Never use the stand for any other purpose. • The battery will get warm during charging, but this is not a failure. Inserting or removing the plug at short intervals (approx.
  • Page 11 • Use the handpiece battery charging stand for the iStar handpiece only. Do not put any other handpiece or object (e.g., wire safety pin, etc.) into the stand. CAUTION • Do not charge, use or store the product in direct sunlight, a vehicle in the sun, near fire or a stove, or other location where temperatures are outside the range shown in the specification table in this manual.
  • Page 12: Maintenance

    5-5 Auto power-off function When approximately 10 minutes have elapsed with no operation of the ON/OFF key the power is automatically turned off to save energy and prevent accidental operation. The power is NOT turned off while the handpiece motor is rotating. 5-6 Last memory function The current setting (last memory) is always memorized when turning off the handpiece motor.
  • Page 13: Periodic Maintenance

    Periodic Maintenance Perform periodic maintenance checks every three months according to the check sheet below. If any abnormalities are found, contact your Authorized StarDental Dealer for repair. Point to check Details Rotate the handpiece motor and check for abnormalities such as abnormal Rotation rotation, vibration, noise, and overheating.
  • Page 14: Troubleshooting

    Troubleshooting When malfunctions occur, check the items below prior to requesting repair services. If no item is applicable or the trouble is not eliminated after following the recommended actions contact your Authorized StarDental Dealer for repair services. <Handpiece motor, stand, AC adapter> Trouble Cause Remedy...
  • Page 15: Classification

    Classification • Type of protection against electric shock: - Class II equipment • Degree of protection against electric shock: - Type B applied part (Applied part: Disposable prophy angle, Handpiece motor) • Method of sterilization or disinfection recommended by the manufacturer: - See "6-1 Cleaning the handpiece motor and stand"...
  • Page 16: Warranty

    ® DentalEz warrants all components of the iStar Cordless Prophy Handpiece for 12 months from the date of purchase*; with (1) free battery replacement for the inner module within the first (3) years. Please note the following additional terms of our warranty policy: •...
  • Page 17: Disposal

    Disposal Dispose of the stand and AC adapter as industrial waste according to local laws and regulations. Dispose of the handpiece motor as medical waste. The handpiece motor includes a battery pack (lithium-ion battery). Do not dispose of in a fire or heat it.
  • Page 18 Guidance and manufacturer's declaration - Electromagnetic Immunity The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment. Immunity test IEC60601 test level Compliance level Electromagnetic environment - guidance Conducted RF...