Ashcroft 1305D Installation And Maintenance Manual

Ashcroft 1305D Installation And Maintenance Manual

Deadweight tester and portable pump
Hide thumbs Also See for 1305D:

Advertisement

1.800.561.8187
www.
.com
Installation and
Installation and
Maintenance
Maintenance
Manual for the
Manual for the
®
ASHCROFT
®
Type
ASHCROFT
Type
1305D Deadweight
1305D Deadweight
Tester and Type
Tester and Type
1327D Portable
1327D Portable
Pump
Pump
I&M002-10095 4/02 (250-1526) Rev. 7/05
information@itm.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 1305D and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Ashcroft 1305D

  • Page 1 Installation and Maintenance Maintenance Manual for the Manual for the ® ASHCROFT ® Type ASHCROFT Type 1305D Deadweight 1305D Deadweight Tester and Type Tester and Type 1327D Portable 1327D Portable Pump Pump I&M002-10095 4/02 (250-1526) Rev. 7/05 1.800.561.8187 information@itm.com www.
  • Page 2 1.800.561.8187 information@itm.com www. .com...
  • Page 3: Purpose And Scope Of Manual

    1.0 PURPOSE AND SCOPE OF MANUAL for the Ashcroft ® Type 1305D Deadweight Tester This manual is provided to guide users of Model 1305D and Type 1327D Portable Pump and 1327D pressure testing devices in: (1) installing the equipment INDEX...
  • Page 4 NPT fitting adapters are supplied. ciable length of time regardless of temperature changes, Operating Fluid: 1305D – Light grade machine oils, auto- slight leaks in the pressure system, or changes in volume motive petroleum base SAE 20 oils or other equivalent flu- of the pressurized system due to movement of a Bourdon ids suitable for use with Buna N O-ring materials.
  • Page 5 Number of Weights by Value Weight L-10 L-20 L-40 L-100 High psi Type High High lbs. H-25 H-50 H-100 H-200 H-500 1305D-10 1305DH-10 15/200 75/1000 1305D-20 1305DH-20 15/400 75/2000 1305D-30 .0625 in. .0125 in. 1305DH-30 15/600 75/3000 (.4032) (.0806) 1305D-50...
  • Page 6: Setting Up The Tester

    Section 1.0. Damaged merchandise and the packaging materials must be held for inspection 2.2.3 Specifications – 1327D by authorized Ashcroft Inc. personnel or the responsible Accuracy: ⁄ % of 1% of test gauge span. carrier. Disposition of damaged or missing goods will be...
  • Page 7 Unit Of Model Tools & Net Weight Measure Number Case Access. Gauge Range(s) Included lbs. 1327* 1327H* – – – – – – 1327O 1327D 1327DH – – – – 1327DO 1327DG-2 1327DGH-2 0/150 – – – 1327DGO-2 1327DG-6 1327DGH-6 0/150 0/600 –...
  • Page 8 3.3.1 Piston and Cylinder Selection attention providing the reservoir level is maintained. The 1305D Tester is a dual range device in that most -models utilize two interchangeable piston and cylinder 3.3.3 Weights...
  • Page 9 Figure 2-1 Ashcroft Portable Deadweight Tester 1305D 1.800.561.8187 information@itm.com www. .com...
  • Page 10 3.3.6 To Dismantle the gauge socket. Rotate the gauge for viewing by loosen- To replace the tester in its case, remove the gauge offset ing nut adapter (62), setting the gauge to the desired posi- pipe assembly (64 or 66) and the piston and cylinder tion, and re-tightening nut adapter.
  • Page 11 Unscrew the gauge from bushing (59) if used, or connec- The mechanism of an Ashcroft Pressure Gauge includes tor (60), and remove. a Bourdon tube and geared movement, which is unaffect- Remove the test gauge in the same manner as the gauge ed by variations in gravity.
  • Page 12: Maintenance Instruction

    When removing or replacing the piston, it should be as with the 1305D Deadweight Tester, a buoyancy correc- rotated back and forth. tion is not necessary. In other designs, the piston is some-...
  • Page 13 Slide check valve (44), check valve spring (14) and Dead Weights valve space (29) into body assembly (2). Piston and Cylinder Assembly – 1305D c. Replace bleed plug (50) with O-ring packing (27) into Test Gauges – 1327D body assembly (2).
  • Page 14 Figure 2-2 Ashcroft Portable Test Pump – Type 1327D 1.800.561.8187 information@itm.com www. .com...
  • Page 15: Shipping Instructions

    The piston/cylinder assembly is available as a unit only. This assures the user of maximum accuracy in his Ashcroft Deadweight Tester and maintains certified trace- ability to the N.I.S.T. 6.0 WARRANTY – SHIPPING INSTRUCTIONS –...
  • Page 16 Reassemble unit. placement is required. *These symptoms apply only to the 1305D Deadweight Testers. High pressure Air trapped in Reprime pump. does not increase pump piston. Operate pump when pumping...
  • Page 17 1.800.561.8187 information@itm.com www. .com...
  • Page 18: Replacement Parts

    D SERIES – BLUE COLOR REPLACEMENT PARTS – PORTABLE DEADWEIGHT TESTER – TYPE 1305D PORTABLE TEST PUMP – TYPE 1327D Qty. Qty. Unit Item Item List Asmy. Part No. Description Asmy. Part No. DescriptionPrice 603D001-01 Body Only, Pump LD227A Bushing2 9.50...
  • Page 19 AVAILABLE WEIGHTS & WEIGHT SETS INDIVIDUAL WEIGHTS Pressure Range Part Number Units High LF199D psig LPF462 psig LAZ199D psig LAZ199 psig LAZ199A psig LAZ199B psig LAZ199C psig PSI WEIGHT SETS Number of Weights Weight Set by High Pressure Value with Part Carrying Number...
  • Page 20 All specifications are subject to change without notice. All sales subject to standard terms and conditions. I&M002-10095 4/02 (250-1526) 1P10/92 8P5/07 ©Ashcroft Inc. 1.800.561.8187 information@itm.com www. .com...