17
Technical Information
Interconnections
It is a 15-pins D type connector; the available signals are detailed in
the table, the following paragraphs describe the signal characteristics
and use.
Table 2
Pin N.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

Signal description

START/STOP: input signal to start or stop the pump. With the
supplied cover connector the START/STOP (+) signal is connected to
the +24 Vdc pin and the START/STOP (-) signal to the GROUND pin:
in this condition the pump automatically starts as soon as the
controller recognises the input supply ("Plug & Pump").
INTERLOCK: input signal to control the pump rotation. With the
supplied cover connector the INTERLOCK (+) signal is connected to
the +24 Vdc pin and the INTERLOCK (-)signal to the GROUND pin.
180/224
Signal characteristics and use
Signal name
Start/Stop (+)
Start/Stop (-)
Interlock (+)
Interlock (-)
Speed setting (+)
Speed setting (-)
Soft Start (+)
Soft Start (-)
+ 24 Vdc
Spare
Programmable set point
Spare
Fault Output
Programmable analog signal (+)
• Ground
• Programmable analog signal (-)
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
In/Output
In
In
In
In
In
In
In
In
Out
Out
Out
Out
Out