This manual may be changed as the version is upgraded, without previous notice. Inquiries If you have any kinds of inquiries for this model, please contact your local agent or Head Office. Head Office : SEWHA CNM CO., LTD. Website : http://www.sewhacnm.co.kr Email : info@sewhacnm.co.kr sales@sewhacnm.co.kr...
7. Item’s performance will be up-dated continuously base on previous version’s performance. 2-3. Features 1. SI 580 model is the standard 1/8 DIN SIZE and compact enough, so it is easy to install. 2. It has wide range of DC Input.
3-2. Front Panel 3-2-1 Front Panel (Display / Key Pad) Weight Display FND 3-2-1. Status Lamp STEADY When the weight is “STEADY”, Lamp is ON. ZERO When the current weight is ”ZERO”, Lamp is ON. TARE “TARE” function is set, Lamp is ON. HOLD “HOLD”...
Page 7
5.Test Mode 1 : Analog value check mode 6.Test Mode 2 : Relay out check mode 7.Set up Mode : Enter F-FUNCTION Mode 1. To set the “HOLD” Function (refer F10) [1 input : “HOLD”, input : “HOLD Reset” ] 2.Calibration Mode : Move to right 3.F-Function setting : Move to right 3.
3-3 Rear Panel 5. INPUT 6.RELAYOUT 7.ANALOGUE OUT 8.SERIAL I/F 1.Power 2.INPUT 3.SERIAL I/F 4. LOAD CELL 1. Power AC IN: 18V~24V (Power : 24V 1A recommended) 2. External Input terminal: Standard tow port (Refer to F-Function F14, F15, F16, F17 to select desired function of each input terminal ) 3.
4-3. Load cell Installation Load Cell Wire Connection (In case of SEWHACNM’s Load cell) It depends on the manufacturer of load cell, please check the specification. ※ Load cell wire color can be changed without prior notice. Under set up the Load cell, if EXC+ and EXC- have a short circuit, It may cause damage in the indicator.(specially analogue board)
5. SET-UP 5-1. Set up This is the Menu which can set the all of the functions. There may be some display differences between real and on the manual. 5-1-1. Start “SET UP” Mode (Pass is Word Not used) PUSH FOUR TIMES “SET UP”...
■ (Calibration) Adjusting “ZERO” Balance Adjust weight balance between “Real weight” on the load cell(Weight Part) and “Displayed weight of Indicator”. When you replace LOAD CELL or Indicator, you have to Calibrate process once again. (When you start calibration mode, TARE, HOLD & PRINT will be reset.) Before processing calibration, please warm up the indicator during 15 min to guarantee more preciseness.
Page 13
5-2-2. Setting “Capacity of weighing Scale” After displaying “CAPA”, input max capacity with keys & Press key to save & move to next step. If you want that set Max capa is 1,000kg, then just input “1000”. You don’t need to consider Division value. 5-2-3.
Page 14
5-2-4. Measure the “DEAD” Weight of Weighing Scale. When “DEAD” displays, press key, then indicator will calculate Dead weight of scale part automatically. Indicator will search “DEAE weight” during 10~20 secs automatically to find the best condition. ※ Over than 1/10,000 resolution setting, To guarantee the preciseness, DEAD weight calculation (CAL00~CAL09) will be operated twice.
Page 15
If “UP” is displayed, please load “Test Weight” on the scale part and press key. Calculate Span value during 10 ~20 secs, automatically. ※ Over than 1/10,000 resolution setting, To guarantee the preciseness, Span calculation will be operated twice. After calculation, span value will be displayed on When “...
5-3. Simulation Calibration Mode(Calibrate without Test weight) With this “Simulation Calibration Mode” you can make simple calibration without any “TEST weight” This calibration mode uses “Load cells’ max capacity” and “Max. Output Rate(mV)”, so the weight adjustment degree might be less than “Test weight Calibration”. The guaranteed resolution of this “Simulation Calibration”...
Page 17
5-3-2. Setting “Capacity of Load Cell” After “CAPA” displayed, Check Max. Capacity of Load cell. (refer the load cell label, or Test Report) Input the Max. Capacity of Load cell. And press key. In case of plural piece of load cells are installed, Please make sum of each load cell’s capacity and make setting with Max.
Page 18
5-3-4. Measure the “DEAD Weight” of Weighing Scale. “dEAd” is displayed. Please press key with empty scale. Then the indicator starts to measure and find optimal “Dead weight value of Scale” automatically. It takes 10sec or 20sec to get the best situation.
Page 19
If input wrong value, there will display “BAD”, please Setting “Capacity of Load Cell”. go back to After recheck the label of load cell and retry the process. After displaying “mV”, input Load cell Output Rate(mV), referring the load cell label. And press key to save.
5-4. F-FUNCTION Setting Set-up means set the F-function and make optimal operation of SI 580 controller. Starting F-FUNCTION Mode After Displaying “SETUP” press key. FUNCTION FUNCTION NUMBER SET VALUE After input set value, press key to save. Whenever press key, function NO. will increase 1 by 1.
Page 21
■ F-FUNCTION LIST General Function Setting (“●” Factory default set value) Weighing Data Save Method selection ● Non-Save Mode (Weight Data & Frequency) Save Mode ( Weight Data & Frequency) Weight –Back up selection Normal Mode ● Weight Back up Mode Motion Band Range setting This is set “Steady”...
Page 22
CAUTION : If setting over than 10%, The display weight could be over than Load cell input signal or Max. Capacity and it may display “CELL-Err” or incorrect weight value. And It can be the cause of load cell damage. Tare key Operation Range selection: (-) value is same to (+) Activated within 10% of Max.
Page 23
External Input Selection 2 (IN2) – Bottom side Terminal Not used HOLD/HOLD RESET ZERO RUN (F21-03 Under Packer Mode) TARE STOP (F21-03 Under Packer Mode) TARE RESET RUN / STOP(F21-03 Under Packer Mode) ● TARE / TARE RESET PRINT HOLD PRINT the amount HOLD RESET External Input Selection 3 (IN3) –...
Page 24
“FINISH Relay” delay time setting (Under F21-03/04 setting) 0.1 sec / Unit Ex)10 : 0.1 x 10 = 1 sec “Finish Relay” “ON” time setting (Under F21-03/04 setting) 0.1 sec / Unit Ex)10 : 0.1 x 10 = 1 sec ■...
Page 25
“Check-Sum” detection selection (Under F32-01 setting, only) ● Check-Sum Not Use Check-Sum Use Under Stream Mode select the way transmit data protocol/frame (basic port) ● Transmit by Protocol Transmit by frame (in case of using specific utility) Caution : In case of “Transmit by frame” & under 14400bps setting(F31), the speed of system will be slow.
Page 26
Paper Withdraw Rate setting (After SUB/GRAND Total Print) Whenever set value increased, 1 line will be added. Paper Withdraw Rate setting (After Continuous/Single Print) Whenever set value increased, 1 line will be added. Printing Language Selection ● KOREAN ENGLISH Minus(-) symbol Print selection ●...
Page 27
Serial Communication Speed selection(Extended Serial port) 2,400bps 28,800bps 4,800bps 38,400bps ● 9,600bps 57,600bps 14,400bps 76,800bps 19,200bps 115,200bps DATA Transference Method selection(Extended Serial port) ● Simplex Mode / Stream Mode Duplex Mode / Command Mode Print Mode “Check-Sum” detection selection (Under F62-01 setting, only) ●...
Page 28
■ Other Setting Mode EMPTY Range setting You can set “EMPTY” Range. Ex) “0” setting : When Net Zero, “Zero” status lamp is ON. “200” setting : Under “200”, “Zero” Status lamp is ON. TIME(H,M,S) Check / Change (every 24Hours) Check Current DATE data or you can Change to new date DATE(Y,M,D) Check / Change Check Current TIME data or you can Change to new time...
Page 29
5-5. SET-POINT Setting (After input Function no. and press “F” key.) (Each Control Relay Set point Value setting- refer 21) There is “SETUP”, then press After input SET-POINT NO., input set value and press key to save.
Page 30
◆ Weighing Mode 1 – Limit Mode 1 (F21 – 01 Setting) – Relay “ON” when weight reaches set value WEIGHT Relay Output Relay Contents Relay Contents Current weight ≥ SP1(ON) Current weight ≥ SP2(ON) OUT 1 OUT 2 Current weight < SP1(OFF) Current weight <...
Page 31
◆ Weighing Mode 2 – Limit Mode 2 (F21 – 02 Setting) – Relay “ON” when weight reaches set value. User’s relay, “A” dry WEIGHT Relay Output Relay Contents Relay Contents current weight ≥ SP1(ON) current weight ≥ SP2(ON) OUT 1 OUT 2 current weight <...
Page 32
◆ Weighing Mode 3 – Packer Mode 1 2 Step control Packer Mode -F21 – 03 Setting – Relay “ON” when weight reaches set value - Relay “ON” Within “EMPTY” range WEIGHT Ex) When input set value as SP1(Target) = 1000, SP2(Drib) = 200, Empty range = 10, SP3(free fall) = 0 Start input : SP1, SP2 and SP4 will be “ON”.
Page 33
◆ Weighing Mode 4 – 3 Steps (F21 – 04 Setting) , Relay “ON” at finish point WEIGHT EX) SP1(Target) = 1000, SP2(Drib) = 200, SP3(Bulk) = 500, SP4(Finish)= 50 Start Input : SP1, SP2,SP3 will be “ON” Current weight will reaches to 500(SP1-SP3), SP3 will be “OFF”. Current weight will reaches to 800(SP1-SP2), SP2 will be “OFF”.
Page 34
◆ Weighing Mode 5 – Packer Mode 3 3 Steps Control Packer Mode (F21 – 05 Setting), Relay “ON” at Empty range WEIGHT EX) SP1(Target) = 1000, SP2(Drib) = 200, SP3(Bulk) = 500, SP4(Finish) = 50, Empty Range=10 Start Input : SP1, SP2, SP4 will be “ON”. Current Weight reaches to Empty range(=10) , SP4 will be “OFF”.
Page 35
◆ Weighing Mode 6 – Limit Mode 3 (F21 – 06 Setting) – Relay “ON” when weight reaches set value. User’s relay, “B” dry WEIGHT Relay Output Relay Contents Relay Contents current weight ≥ SP1(ON) current weight ≥ SP2(ON) OUT 1 OUT 2 current weight <...
Page 36
◆ Weighing Mode 7 – Accumulating Mode 1 (F21 – 07 Setting) WEIGHT Relay Output Relay Output Relay Outut Current weight ≥ SP1(ON) Current weight ≥ SP2(ON) OUT 1 OUT 2 Current weight < SP1(OFF) Current weight < SP2(OFF) At SP3 Current weight ≥...
Page 37
◆ Weighing Mode 8 – Accumulating Mode 2 (F21 – 08 Setting) WEIGHT Relay Output Relay Output Relay Outut Empty<Current weight < SP1(ON) SP1<Current weight < SP2 OUT 1 OUT 2 Current weight ≥ SP1(OFF) Current weight ≥ SP2(OFF) At SP4 SP2<Current weight <...
5-6. Test Mode Before starting the TEST mode, please remove other connected devices. Under “SETUP” mode, press key. TEST MODE 1 TEST MODE 2 TEST MODE 1 ESC / BACK Analog value Analog Key/Digital Input Go back to Check Mode Variation Value Check Mode Test Mode 2...
Page 39
A/D board 5-6-1. Test Mode 1) Analog Check Mode Under this mode, you can check analogue value to real digital value through Display. The last digital value can be fluctuated. Displaying 1~100,000 Displaying 10~1,000,000 Displaying 100~10,000,000 2)Analogue Value Check Mode Under this mode, you can check the variation degree of analogue value.
Page 40
3) Key / Digital input Test Mode Under this mode, you can test Key input and Digital Key input test. First display position is for key Second display position is for pad input digital input Whenever pressing key pad or plus to digital input terminal, the matched No. will be displayed on the each position.
Page 41
6) Standard Serial Interface Test Mode. Connect with PC or other devices through serial interface and check the transference and receipt. At the normal operation, display will be blinked. 7) Extended Serial Interface Test Mode. Connect with PC or other devices through serial interface and check the transference and receipt. At the normal operation, display will be blinked.
Page 42
6. INTERFACE 6-1-1. Serial Interface (RS – 422) : Standard (selectable) 32 pcs of Indicators (SI580) are connectable .(SI480, SI580) (SI580) 6-1-2. Serial Interface (RS – 485) : Standard installed (selectable) 32 pcs of Indicators (SI580) are connectable (SI480, SI580)
Page 43
6-1-3. Serial Interface (RS – 232) GND(T+) Serial communication interface is sensitive to electric noise. Install isolated place from Power cable or other electric cables and wires, and please use shielded cable for better performance.
6-1-4. Data Format 1. Data Format1 : ID Number is not be transferred.(Refer “FUNCTION 37/67-00”) Header1 Header2 OL : OVER LOAD NT : NET-WEIGHT(Tare is not set) ST : Stable GS : when setting TARE US : Unstable 2. Data Format2 : ID Number + Data Transference (Refer F-function 37/67-01, F18) Header1 Header2 OL : OVER LOAD...
Page 45
4. CAS Format (22byte) (Refer F-function 37/67-03) LAMP DISPLAY Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 Gross STABLE Hold Print TARE ZERO Weight Header1 Header2 OL : OVER LOAD NT : GROSS weight ST : Stable GS : Net weight US : Unstable...
Page 46
HEX : 02 30 31 52 43 57 54 03 STX ID RCWT State1(1byte) State2(1byte) P decimal point(1byte) +/-(1byte) Current weight(7byte) Weight unit(2byte) ETX SI 580 response State1 : O(over weight) , S(Steady), U(Unsteady) State2 : N(Net weight), G(Gross weight) Ex) Steady(S), TARE not used(N), 0.000kg...
Page 47
Current Time Data ASCII : STX ID(2Byte) RTIM ETX HEX: 02 30 31 52 54 49 4D 03 STX ID RTIM Current Time(6byte) ETX SI 580 response Ex) Time : 12:00:00 Current Date Data ASCII : STX ID(2Byte) RDAT ETX...
Page 48
ASCII : STX ID(2Byte) RSP3 ETX HEX : 02 30 31 52 53 50 33 03 STX ID RSP3 P decimal point(1byte) set value(7byte) ETX SI 580 response decimal point Ex) SP3 set value : 7.000 Set Point4 set value...
Page 49
Set as Zero (same as “ZERO” key) ASCII : STX ID(2Byte) WZER ETX HEX: 02 30 31 57 5A 45 52 03 SI 580 response normal: STX ID ACK ETX error: STX ID NAK ETX TARE ASCII : STX ID(2Byte) WTAR ETX...
Page 50
DATE Setting ASCII : STX ID(2Byte) WDAT current DATE (6byte) ETX Ex) Date : Aug 12 ,2009 SI 580 response normal: STX ID ACK ETX error: STX ID NAK ETX TIME Setting (2Byte) ASCII : STX ID WTIM Time (6byte)ETX...
Page 51
Setting SP2 (2Byte) ASCII : STX ID WSP2 SP2set value(7byte) ETX Ex) SP2 : 6.000kg (decimal point : 0.000) SI 580 response normal: STX ID ACK ETX error: STX ID NAK ETX Setting SP3 (2Byte) ASCII : STX ID WSP3 SP3set value(7byte) ETX Ex) SP3 : 7.000kg (decimal point : 0.000)
Page 52
■ Command Mode Example READ COMMAND Ex.) Current Weight Command(RCWT), ID No. : 01, Current Weight : 1,000kg 1) P.C Read Command Format (STX ID NO. RCWT ETX) “Check-sum” not used. PC transmits to SI580 SI580 Response to PC 2) When PC requests to Indicator, Format(STX ID RCWT ETX) CHCEK SUM is used. PC transmits to SI580 SI580 Response to PC...
Page 53
■ WRITE COMMAND Ex) SP1 Setting Command, ID No : 01, New SP1 Set value : 0.600kg 1) PC Write command format (STX ID WPR1 000.600 ETX) “CHECK SUM” not use. PC transmits to SI580 SI580 Response to PC Normal operation Incorrect operation 1) PC Write command format (STX ID WPR1 000.600 ETX) “CHECK SUM”...
Page 54
6-2. Relay Output 4pcs Control Relay output is installed in Output Terminal. You can make setting for these relay output through F21 function.. (A/B Dry Contact). Relay Specification 12VDC Coiling Rating 1A 24VDC Contact Ratings Relay Output Diagram. Under TEST Mode ,Calibration mode and SET-POINT setting mode, the relay output will be OFF.
Page 55
6-3. Analogue Output Interface. (4~20mA : Factory default set value This output card converts weight value to Analog output signal (4~20mA) and transfers to external devices(Recorder, P.L.C), controlled by voltage output. 6-3-1. Specification Output Current 4~20㎃ (Output Range :2~22㎃) Accuracy More than 1/1,000 Temperature Coefficient 0.01%℃...
Page 56
6-4. Analog Output Interface (0~10V) This output card converts weight value to Analog output signal (0~10V) and transfers to external devices(Recorder, P.L.C), controlled by voltage output. 6-4-1. Specification Output Voltage 0~10V DC output Accuracy More than 1/1,000 This output is adjusted as when the weight is “Zero”, output is 0V and when the weight is “Full capacity”, output is 10V.
Page 57
6-5 Serial Print (F32/F62-02 setting) – RS-232 Serial Interface. It can be connected with all kinds of Serial interface printer, but the printing format is already programmed and fixed with SE7200/7300 model. Printing Format Using the RS-485 or 422 interface, please use convertor and converts to RS-232 and connect with Serial printer.
Page 58
7. Error & Treatment 7-1. Load Cell Installation Error Cause Treatment Remarks 1. Load cell broken 2. Load cell isolation 1. Input Resistance of “EXC+” and “EXC-“ is resistance error 1. Measure about 400Ω. ±3 3. Weighing part input/output resistance Weight Value is 2.
Page 59
7-2. Calibration Process Error Cause Treatment When Max.capacity/digit value is over Re-input the Max. Capacity, less than 20.00 Err 01 20,000 (Max. Capacity / Digit) Standard weight value is over than Re-input Standard weight value with Number Err 04 Max. Capacity keys, under Max.
Page 60
7-3. Digital Weighing Indicator Error Display Cause Treatment 1. Under “TEST” mode 1, check 1. Load cell Error analogue value. If you cannot get 2. Load cell cable Error any analogue value or there is no change although adding load, 3.Load cell connection Error please check load cell, load cell 4.
Page 61
WARRANTEE CETIFICATION This product is passed “Sewhacnm’s strict quality test. If there is defect of manufacturing or abnormal detection within warrantee period, please contact our Agent or Distributor with this Warrantee certificate. Then, we will repair or replace free of charge.