This document is believed to be complete and accurate at the time of publication. In no event shall Dionex be liable for incidental or consequential damages in connection with or arising from the use of this document.
1.3 Typical System Setups 1.3.1 Single-Stack In many cases, bench space is limited; hence the Dionex UltiMate 3000 systems are designed to consume as little space as possible. We assume that a single-stack setup is the most common case; therefore, the LCi capillary kits have been optimized for this scenario.
Page 8
Opening and closing Viper connections at high system pressures may reduce the life- time of the fitting system. If you are unable to get a Viper connection tight, call your local Dionex representative or replace it with a new capillary.
In addition, dedicated Viper capillary RS and SD kits containing 3 capillaries each are available to equip Dionex UltiMate 3000 systems with Viper. Figure 6: Packaging box for Viper RS and SD capillary kits for UltiMate 3000 systems (single system, see Figure 1).
The 6-port valve has a direct fluidic connection between port 1 and port 6. To achieve this fluidic connection for a 10-port valve, Dionex recommends connecting ports 7 and 10 with an additional suitable capillary. The inner diameter of this capillary should be 0.13 mm (0.005″) for a rapid separation (RS) setup or 0.18 mm (0.007″) for a standard (SD) setup.
2-position 10-port valve for column switching, a split loop well-plate autosampler, and a stackable Dionex detector, all of which are fully controlled by the Chromeleon Chromatography Data System. Prior to installation, please consider chapter 1 for important additional information.
2.3 Stacking the Modules Please stack the modules as shown below. The picture is a schematic to illustrate the course of the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside the modules as needed. SRD-3600 Solvent Rack and Degasser...
Page 20
In case of a dual-stack setup, the TCC 10-port valve right turns into a TCC 10-port valve left. Both stacks should have a distance of about 7 cm (2.8 inches). Connecting a post-column cooler with a stackable Dionex detector does not work without additional capillaries when setting up the system in the U-configuration (1.3.2).
(right) and one 2-position 6-port switching valve (left) for column switching, a split loop well-plate autosampler, and a stackable Dionex detector, all of which are fully controlled by the Chromeleon Chromatography Data System. Prior to installation, please consider chapter 1 for important additional information.
3.3 Stacking the Modules Please stack the modules as shown below. The picture is a schematic to illustrate the course of the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside the modules as needed. SRD-3600 Solvent Rack and Degasser...
Page 24
350 mm to 550 mm. Connecting a post-column cooler with a stackable Dionex detector does not work without additional capillaries when setting up the system in the U-configuration (1.3.2).
2-position 6-port switching valve for column switching, a split loop well-plate autosampler, and a stackable Dionex detector, all of which are fully controlled by the Chromeleon Chromatography Data System. Prior to installation, please consider chapter 1 for important additional information.
4.3 Stacking the Modules Please stack the modules as shown below. The picture is a schematic to illustrate the course of the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside the modules as needed. SRD-3600 Solvent Rack and Degasser...
Page 28
In case of a dual-stack setup, the TCC 6-port valve right turns into a TCC 6-port valve left. Both stacks should have a distance of about 7 cm (2.8 inches). Connecting a post-column cooler with a stackable Dionex detector does not work without additional capillaries when setting up the system in the U-configuration (1.3.2).
2-position 6-port switching valve for column switching, a split loop well-plate autosampler, and two stackable Dionex detectors, all of which are fully controlled by the Chromeleon Chromatography Data System. Prior to installation, please consider chapter 1 for important additional information.
Quick Installation Guides 5.3 Stacking the Modules Please stack the modules as shown below. The picture is a schematic to illustrate the course of the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside the modules as needed. SRD-3600 Solvent...
6.3.1 Single-stack without Extension Kit Please stack the modules as shown below. The picture is a schematic to illustrate the course of the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside the modules as needed. SRD-3600 Solvent Rack and Degasser...
6.3.2 Dual-stack with Extension Kit Please stack the modules as shown below. The picture is a schematic to illustrate the course of the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside the modules as needed. For the use of the Method Scouting Extension Kit, a dual-stack is recommended. In combination with the second solvent rack, this setup offers space for up to 9 additional solvent bottles and the solvent selection valve to be placed on top of the system.