Health Connector zacurate CMS 500DL User Manual

Finger pulse oximeter

Advertisement

Zacurate® CMS 500DL
Finger Pulse Oximeter
User Manual
Distributed by:
Health Connector LLC
4910 Wright Road
Suite 120
Health Connector LLC
Stafford, TX, 77477
4910 Wright Road, Suite 120
Phone: +1-832-939-8290
Stafford, TX, 77479
http://www.medshopandbeyond.com
http://www.medshopandbeyond.com
orders@medshopandbeyond.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the zacurate CMS 500DL and is the answer not in the manual?

Questions and answers

Summary of Contents for Health Connector zacurate CMS 500DL

  • Page 1 Zacurate® CMS 500DL Finger Pulse Oximeter User Manual Distributed by: Health Connector LLC 4910 Wright Road Suite 120 Health Connector LLC Stafford, TX, 77477 4910 Wright Road, Suite 120 Phone: +1-832-939-8290 Stafford, TX, 77479 http://www.medshopandbeyond.com http://www.medshopandbeyond.com orders@medshopandbeyond.com...
  • Page 2 Copyright © Health Connector LLC Storage and Transportation All rights reserved. Health Connector LLC owns and reserves all of the rights comprised in the copyright of this document. No part of this document may be changed, excerpted, copied, reproduced, or imitated in any form or by any means without prior written consent of Health Connector LLC.
  • Page 3: Table Of Contents

    precise for the combination of the mainframe and probe of the Contents Letter to Users ................1 pulse Oximeter. If the functional tester can measure the portion of Standard Packing List ..............1 overall errors of the mainframe-probe system of the Oximeter Chapter 1 Precautions, Warnings and Symbol ......
  • Page 4: Overall Errors Of The Mainframe-Probe System Of The Oximeter Chapter

    Letter to Users : measured value of pulse oxygen saturation during the first measurement using the finger pulse Oximeter Thank you for choosing our Zacurate® CMS 500DL Finger Pulse : reference value of pulse oxygen saturation during the i Oximeter. By listening intently and applying some of the changes as measurement using the carbon-monoxide-blood-gas analyzer suggested by our customers, we hope you will find that the CMS 5.2.2.3 Characteristics of Population under Clinical Research...
  • Page 5: Precautions, Warnings And Symbol

    measured values of a carbon-monoxide-blood-gas analyzer. Standard Packing List Pulse oximeter………………………….1 PC 5.2.2 Determination of Oxygen Saturation Accuracy AAA Battery……………………………...2 PCS The claimed oxygen saturation is supported by coverage of the entire Lanyard………………………………….….1 PC range of clinical research measurements. Silicon Cover……………………………..1 PC User Manual……………………………...1 PC 5.2.2.1 Data Collection In the clinical test process, data points are recorded with comparable...
  • Page 6 malfunctioning. Oxygen saturation The pulse oximeter uses infrared light (invisible to your eyes)  Resolution to measure your SpO level. Hence, please do not stare at Pulse rate 1 bpm the light-emitting components of the Oximeter to avoid Oxygen ± 2% (70% to 100%) potential eye damage and/or blindness.
  • Page 7: Warnings

    ◆Ambient light intensity is too bright. Hence, please avoid direct exposure to strong light (such as beams from Working Temperature 5C to 40C (41F to 104F) operating lamps or sunlight) during measurement. ◆The probe of the Oximeter is placed on the same arm that a Relative Humidity 15% to 80%, non-condensing blood pressure cuff arterial duct or intravenous injection is...
  • Page 8 Chapter 5 Technical Description and Safety Type Warning: Keep unit and lanyard away from children as the included lanyard may pose as an entanglement or choking Technical Specifications hazard to small children. Adult supervision required; never 1. Dimensions: 58.0 mm (Width) × 32.0 mm (Depth) × 33.2 mm leave children unattended with unit or lanyard (Height) (Approximately 2.28 in X 1.26 in X 1.31 in) Weight: 49.5 g/1.75 oz (including the weight of two AAA batteries)
  • Page 9 Caution: Do not operate the Oximeter if it is wet. Avoid 2) Avoid making any body movement, especially your finger while moving the oximeter from a cold to a hot and humid taking measurement. environment. 3) Long fingernails obstruct the light sensor and prevent accurate measurement.
  • Page 10: Symbol

    Problems Symbol Possible Cause Solution damaged. 3. Contact the distributor. Symbol Meaning The screen 1. If the Oximeter 1. This is normal. turns off does not detect Just turn on the BF-type application part suddenly. any signal within pulse oximeter 16 seconds, it again.
  • Page 11 Problems Possible Cause Solution Symbol Meaning The displayed 1. The finger is not 1. Make sure that value of the placed between the finger is Temperature limitation oxygen the sensor and placed right in saturation the Light between the level or pulse Emitting Diode.
  • Page 12: Chapter 2 About This Product

    Chapter 2 About This Product Troubleshooting Overview Problems Possible Cause Solution stands for peripheral capillary oxygen saturation. Oxygen The Oximeter 1. The finger is not 1. Make sure that saturation is defined as the ratio of oxyhemoglobin (HbO ) to the total fails to display placed between the finger is...
  • Page 13: Product Features

    uses a proprietary algorithm to collect data from pulsatile arterial blood and excludes local noise from the tissues. The relative absorption of light by oxyhemoglobin (HbO) and deoxyhemoglobin is Chapter 4 Maintenance and Troubleshooting then calculated according to the Beer-Lambert's law and a Maintenance quantitative measurement of the users’...
  • Page 14: Limitations

    Limitations during cleaning. Spot check pulse oximeter takes 1-2 readings every second and takes an average of 4-6 readings before displaying the result. Hence, you will have to wait at least 4-6 seconds before a result will be displayed Attention: Do not immerse any part of the instrument into on the monitor.
  • Page 15 1) Increasing the sensitivity of the sensor to allow user to measure Chapter 3 Cleaning and Disinfection his/her SpO levels and Pulse Rate (PR) at blood perfusion as low as Cleaning 0.2%. To clean the instrument, ensure that power is off and remove the 2) Using a self-adjusting smart spring system to fit user’s finger snugly batteries.
  • Page 16: Schematic Structural Diagram

    Product Structure and Operation Steps hypoallergenic. 2.8.4 Reading Oxygen Saturation level and Pulse Rate from the LED Schematic Structural Diagram Screen LED screen Do not move your finger during measurement. Keep your body as still as possible. Once the reading stabilizes, read the Power-on key measured values of the oxygen saturation level and pulse rate on the LED screen.
  • Page 17: Schematic Diagram Of Display

    Schematic Diagram of Display The following figure shows the information display on the LED screen If you do not insert your finger completely into the chamber, of the Oximeter in normal detection state: measurement will be inaccurate. Operation Steps 2.8.1 Installing the Batteries Install the two AAA dry batteries into the battery slot according to the indicated polarity and mount the battery...
  • Page 18 lanyard through the thin end of the lanyard, and tighten the lanyard. Put the silicon cover back on. Ensure that the batteries are installed with correct polarities to avoid damage to the instrument. 2.8.3 Switching On the Power Supply of the Pulse Oximeter Insert one finger completely into the silicon padded finger chamber of the Oximeter.