The essential performance of the device is free from the production of unwanted or excessive stimulation output. INDICATIONS FOR USE The AccuRelief Ultimate Foot Circulator is intended to stimulate muscles of the lower extremity in order to temporarily increase local blood circulation and to facilitate and improve muscle performance.
TABLE OF CONTENTS INTRODUCTION ..................2 INDICATIONS FOR USE ................2 IMPORTANT SAFETY PRECAUTIONS AND WARNINGS ......4 EXPLANATION OF THE ACCURELIEF ULTIMATE FOOT CIRCULATOR ..8 FEATURES ....................9 ADAPTOR ....................10 BATTERY ....................10 DIRECTIONS FOR USE ................12 Foot and Leg Stimulation ( ) ............12 Body Stimulation ( ) .................14...
Page 6
CAUTION WARNINGS AND PRECAUTIONS REGARDING THE PADS • Do not bend or fold because the pad may not function properly. Place the pads onto the plastic film and then store into the sealed package when not in use. • Do not apply ointment or any solvent to the pads or to your skin because it will disrupt the pads from functioning properly.
Page 7
• Replace the cord when broken or damaged. • Do not throw the batteries into a fire. The batteries may explode. • Dispose of the device, batteries, and components according to applicable legal regulations. Unlawful disposal may cause environmental pollution. •...
FEATURES 1) Three output channel electrical stimulator—(1) channel for foot stimulation, (2) channel for body stimulation 2) Power supply—adaptor (included) or 4-C batteries (optional). 3) Settings can be controlled on the panel or with the remote control. 4) The LED displays: treatment time, intensity of foot stimulation and intensity of body stimulation separately.
Page 11
Disposal of batteries Depleted batteries do not belong in the household waste. Dispose of the batteries according to your federal, state and local regulations. As a consumer, you are obligated by law to return depleted batteries. CAUTION 1) Keep the batteries and the product out of the range of children. 2) If the batteries leak, avoid contact with skin, eyes and mucus membranes.
It does not matter if your feet are smaller or bigger than the footpads, as long as pressure is on the instep. Both the footboards of the AccuRelief Ultimate Foot Circulator can tilt back and forth. This gentle and involuntarily feature creates ankle joint movement for increased circulation.
Page 13
• There are 15 therapeutic phases working throughout the cycle time changing every minute. (See page 18 for phase programs) • When using the AccuRelief Ultimate Foot Circulator, the footboards can tilt back and forth as your muscles contract and relax for increased therapeutic value.
Make sure there are no bare metal pins exposed. Step 2 - Insert lead wires to the device. Hold the lead wires plug and insert into the socket on the back of the AccuRelief Ultimate Foot Circulator. CAUTION • Ensure that your AccuRelief Ultimate Foot Circulator is turned OFF.
Page 15
Step 6 - Start treatment and adjust intensity x10000 x100 Using either the remote control or buttons on POWER COUNTER the device, press the -/+ [ ] button to set the intensity level. The treatment will begin and the ] display will start counting down the remaining minutes of treatment.
Stepper Function Tilt your feet back and forth during the treatment. You can press [ ] to shift the display of modes between Stepper and Stimulation. When the STEP LED is lit ( ), the LCD displays your # of steps. When the STIM LED is lit ), it displays treatment time and intensity for stimulation.
SPECIFICATION • Channels: Three, Isolated, (One) for Foot Stimulation, (Two) for Body Stimulation) • Output Amplitude: 0~80mA (500 Ω Load) • Waveform type: Symmetrical Biphasic square pulse with zero DC current • Power Sources: Adaptor (included) or batteries (optional) Input 100V~240V, 50/60Hz Output: DC6V, 300mA Or DC 6.0V, (4) C Batteries (Optional) •...
Program A. Foot stimulation program Phase Frequency Cycle Number (Hz) time time Time 1) Time: 1~99 minutes 2) Frequency: 12.5 10Hz~36Hz 3) Pulse width: 250us 4) On time: 1.9s~10.6s 5) Off time: 1.0s~3.7s 6) Cycle Time: 16.6 3.5s~11.6s 12.5 10.2 10.6 11.6 B.
ELECTRODE PLACEMENT GUIDE Stimulation Position: • Position yourself comfortably in the position as shown in the pictures. The muscles work in an isometric fashion and in a stretched position. To work comfortably and safely, keep your limbs in a fixed position (keep your hands or feet to one spot).
FAQ’S How long is the device warranty? One year. How many times can the electrodes be used? Depending on your skin type, between 10-15 times. Fifteen applications may be possible on skin that is clean, non-greasy and free of excess hair. You should replace the pads when they no longer adhere to the skin or when the muscle contractions are noticeably weaker compared to previous usage.
Page 22
How long should I use the device? The AccuRelief Ultimate Foot Circulator is intended to stimulate muscles to improve or facilitate muscle performance. The device can be used 2-3 times a day for up to 30-60 minutes.
CLEANING AND STORAGE Cleaning the unit 1) Turn unit off and disconnect the lead wires from the unit. 2) Clean the device after use with a soft, slightly moistened cloth and wipe gently. • Do not use chemicals (like thinner, benzene). •...
Storing the electrode pads and lead wires 1) Turn the device off and remove the lead wires from the unit. 2) Remove the pad from your body and pull out lead wires from the pads. 3) Place the pads onto the plastic film and then store into the sealed package.
TROUBLESHOOTING Problem Possible Causes Possible Solution The unit cannot Are the batteries Replace the batteries. power on. depleted? Are the batteries Insert the batteries observing polarity. installed correctly? Is the adaptor con- Reconnect the adaptor to the device nected to the device and main socket according to “Steps and main socket to connect the adaptor”...
Page 26
Problem Possible Causes Possible Solution Stimulation is Intensity is too high. Decrease intensity. uncomfortable Pads are too close. Reposition the pads at least 1 ½” apart. Pad active area size is Replace pads with an active area no too small. less than 2”...
WARRANTY Please contact Carex Health Brands at 800-328-2935 in case of a claim under the warranty. If you have to send in the unit, enclose a copy of your receipt and state what the defect is. The following warranty terms apply: 1) The warranty period for device is one year from date of purchase.
IMPORTANT INFORMATION REGARDING ELECTRO MAGNETIC COMPATIBILITY (EMC) This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
Page 30
Guidance and manufacture’s declaration – electromagnetic immunity The ELECTRICAL STIMULATOR is intended for use in the electromagnetic environ- ment specified below. The customer or the user of ELECTRICAL STIMULATOR should assure that it is used in such an environment. Electromagnetic IEC 60601 Compliance environment –...