11. TROUBLESHOOTING (2)
PROBLEM
LCD Display shows
"EE"
LCD Display shows
abnormal result
No response when you
press button or load
battery
POSSIBLE CAUSE
The cuff was not properly
applied or the rubber
tube was bent or
pressed.
The cuff position was not
correct or it was not
properly tightened
Body posture was not
correct during testing
Speaking, arm or body
movement, angry,
excited or nervous during
testing
Incorrect operation, or
strong electromagnetic
interference
10/15
SOLUTION
Review the cuff applying and testing
sections of the instructions and
re-test.
Apply the cuff correctly and try again
Review the body posture and testing
sections of the instructions and
re-test.
Re-test when calm and without
speaking or moving during the test
Take out batteries for five minutes,
and then reinstall all batteries.