RF A 300 Instruction Manual page 12

Broadband power amplifier

Advertisement

d)
e)
f)
g)
h)
i)
4.4.2
Measurement of Harmonics
Sweep Generator
10 KHz – 35 MHz
RF Output
Figure 4-2.
1.) Equipment Required:
a)
Sweep/Signal Generator
b)
Calorimetric Power Meter HP435B
The material contained in this document is the property of Electronics & Innovation Ltd., it is
subject to change without notice.
Ocotber 2008
Adjust the output level of the sweep/generator for full vertical
deflection on the oscilloscope face.
Calibrate the scope face to show 3 dB in 1 dB steps by
attenuating the sweep/generator in 1 dB.
Return sweep/generator output level to full deflection. Rotate the
step attenuator (CCW) so that the output is reduced by 50 dB.
Reconnect the A300 into the test set-up of Figure 4-1.
Place the A300 power switch to the "ON" position.
Observe the gain versus frequency sweep on the oscilloscope.
The average gain should be greater 55 dB
The gain variation should be within the 3 dB as shown on the
oscilloscope.
Power Meter
30 dB Attenuator
30 dB
coupler
Spectrum
12
Revision D