If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our worldwide contact list for a Hanna representative near you at www.hannainst.com. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket,...
Page 3
If Each instrument is supplied with: service is required, contact your local Hanna O ce. If under warranty, report the model number, date of purchase, serial number and the nature of the problem.
5000 μS/cm, 20 mL sachets (25 pcs.) measurement con guration and is useful for many routine applications. (Every HI 7030M 12880 μS/cm, 230 mL bottle feature and measurement detail is designed to give you an edge in measurement HI 7031M 1413 μS/cm, 230 mL bottle technology.) HI 7033M 84 μS/cm, 230 mL bottle...
Weight 250g (8.82 oz) Liquid Crystal Display (LCD) Micro USB device connection for Capacitive Touch Keypad power or PC interface 3 mm jack input for edge ® Standard USB host connection digital probes for data transfer to a USB Top mounted ON/OFF button...
Probe Speci cations Salinity Diagram 0.00 to 29.99 μS/cm, 0.00 to 14.99 ppm (mg/L), 30.0 to 299.9 μS/cm, 15.0 to 149.9 ppm (mg/L), 300 to 2999 μS/cm, 150. to 1499. ppm (mg/L), 0.0 to 400.0% NaCl *, 3.00 to 29.99 mS/cm, 1.50 to 14.99 g/L, Range 2.00 to 42.00 PSU *,...
Check the keyboard or contact your displays all LCD tags local Hanna O ce. permanently. CAL/MODIFY - Used to enter - Used to scroll through CAL “Prod” message at Instrument was not Contact Hanna Technical Support for exit calibration mode. SETUP menu. Used change startup.
Maintenance Guide to Rinse the probe with clean water after measurements. If a more thorough cleaning is required, remove the probe sleeve and clean the probe with a cloth Indicators EC Probe or a nonabrasive detergent. Make sure to reinsert the sleeve onto the probe Maintenance properly and in the right direction.
Salinity The main operating modes of edge ® are setup, calibration, measurement, Setup/ Measurements If the reading is out of range, the full-scale value data logging, and data export. Follow this general outline of steps to get you Installation (400.0%) will be displayed blinking.
= -7.0261 = 0.0636 Electrode = 2.7081 = -0.01442 Connect the 3 mm probe jack to the probe & Probe input located on the bottom of edge ®. Connections T-15 Make sure the probe is completely f(t)= 1 + 0.0162•(t-15) connected.
*Only seen natural water samples). charging battery when cable (and meter connection LOG ON EDGE or will be used LOG ON EDGE Available To change the temperature coe cient, enter the SETUP mode and select between micro EXPORT TO PC...
Page 12
General NaCl % GLP Option Description Choices Default Basic mode If enabled, the number of days since the calibration Setup Information Used to save expired. (i.e. “CAL EXPIRED 2 DAYS AGO”). battery life by automatically turning o when 5, 10, 30, 60 no key press is Set Auto O 10 min...
• If powering edge ® through the micro USB connector to a PC, a SETUP option will require the choice “LOG ON EDGE” or “EXPORT TO PC“. • 1000 log records can be stored into edge ® memory. This memory is shared between all logging types (Manual, Manual Stability, Interval logs).
Page 14
Logging In Setup mode, choose log parameter, press MODIFY then use the u arrow to EC/TDS GLP select between Interval, Manual, or Stability. When Interval is displayed, use the Function The reference temperature together with the current Information p and q arrows to select the setting for the timed interval. When Stability is reading.
Use the pq keys to scroll through the calibration data displayed on the third LCD line. The cell factor in cm determined from the calibration All log records stored on edge ® may be viewed on the meter by pressing the Viewing with the current reading.
Viewing NaCl % Preparation Delete Logging Type/Lot Logged Pour a small quantity of the calibration solution into a beaker. If possible, use a Calibration Press RCL and select the parameter log. Data plastic beaker to minimize any EMC interferences. Use the pq keys to select the Manual/Stability Before pressing CAL verify in SETUP: records or Interval lots to delete.
Page 17
“CLEAR ALL” will be displayed with “CFM” and “EC“ Conductivity vs Temperature Chart tags blinking. Press CFM. The following table lists the temperature dependence of Hanna EC calibration standards. edge ® uses these values and their temperature coe cients during calibration.
Select standard to be used for calibration. 0.00 μS (probe in air) may be used for measurements. an o set calibration. This should be done rst. Hanna Conductivity standards are available at 84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm and are used for the cell factor calibration.
Turn on edge ®. conductivity 200.0 mS/cm, range and Press SETUP and select “LOG ON EDGE”. 500.0 mS/cm unit. If a xed Press MODIFY then use pq keys to change to “EXPORT TO USB”. value is used, Press CFM and the USB/PC Tag is displayed.
Page 20
User may 0.01 to 9.999 1.000 cm Not available. The four-ring EC probe may be used for 3 di erent measurement applications enter actual Automatically with edge ®. Cell factor determined value Manual during • It may be used for temperature compensated or absolute conductivity calibration.