Read before using
QUATTRO EMS
Model: GM380E (4CH EMS)
FDA 510k
Operation Manual

Advertisement

Table of Contents
loading

Summary of Contents for QUATTRO EMS GM380E

  • Page 1 Powered Muscle Stimulator (EMS) Read before using QUATTRO EMS Model: GM380E (4CH EMS) FDA 510k Operation Manual...
  • Page 2: Table Of Contents

    TABLE CONTENTS GENERAL DESCRIPTION SYSTEM COMPONENTS WARRANTY INDICATIONS AND CONTRAINDICATIONS WARNINGS AND PRECAUTIONS ABOUT THE DEVICE THE DEVICE CONTROLS DANGER ATTACHING THE LEAD WIRES ELECTRODE SELECTION AND CARE TIPS FOR SKIN CARE CONNECTING THE DEVICE BATTERY INFORMATION CHANGING THE BATTERY 9-VOLT DC ADAPTOR CLEANING FOR YOUR DEVICE TROUBLESHOOTING...
  • Page 3: General Description

    GENERAL DESCRIPTION Electric stimulation has proven its high value as a method of pain therapy and is a great help to the experienced therapist. With some indications, physicians can prescribe a unit to patients for the use at home. This series of device is the combination of two units with dual-Channel electric stimulator for active treatment application, which is equipped with a Liquid Crystal Display indicating operation modes and output as well as an 8-bit micro computer for controlling the system.
  • Page 4: Warnings And Precautions

    Indications for use Relaxation of muscle spasms. Prevention or retardation of disuse atrophy. Increasing local blood circulation. Muscle re-education. Maintaining or increasing range if motion. Muscle re-education. Immediate post-surgical stimulation of calf muscle to prevent venous thrombosis. Contraindications Powered muscle stimulators should not be used on patients with cardiac demand pacemakers.
  • Page 5: About The Device

    c. Over the menstruating or pregnant uterus; and d. Over areas of the skin which lack normal sensation. Some patients may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium. The irritation can usually be reduced by using an alternate conductive medium, or alternate electrode placement.
  • Page 6: The Device Controls

    THE DEVICE CONTROLS Intensity The intensity knobs are located on the front of the unit for the strength adjustment of the stimulation and also function as ON/OFF controls. Mode The Mode key is used to select / set the type of treatment utilized. The available stimulation modes for each unit are as indicated in the following table: Identification Function as...
  • Page 7: Electrode Selection And Care

    The lead wires provided with the device insert into the jack sockets located on rear of the unit. Holding the insulated portion of the connector, push the plug end of the wire into one of the jacks. After connecting the wires to the stimulator, attach each wire to an electrode.
  • Page 8: Battery Information

    with your electrodes. Before attaching the electrodes, identify the area in which your physician/practitioner has recommended for electrode placement. 2. Connect lead wires to the electrodes Connect the lead wires to the electrodes before applying the electrodes to the skin. Note: Be sure all intensity controls for every Channel 1 and 2 of each unit are turned to the “OFF”...
  • Page 9: 9-Volt Dc Adaptor

    9-Volt DC Adaptor Because the current consumption is large for the device, it is recommended to use the 9V DC adaptor. Plug the DC plug into the DC jack of the unit, and plug the adaptor into your 110V wall outlet. Please be sure you are using the correct polarity of DC plug.
  • Page 10: Technical Specifications

    TECHNICAL SPECIFICATIONS Channel: Four, isolated between channels Pulse amplitude: Adjustable 0-100mA peak into 500 Ω load each channel, constant current Pulse Rate: 1Hz-160Hz (adjustable), 1Hz/step Pulse Width: 300µs fixed. Software ramp up feature: Pulse width ramp up when change mode Patient Compliance Shows the treatment times.
  • Page 11 Remark: Printed in Taiwan OP-380E-Q01...