  • Page 1 BetaProbe TI/TI+ High Accuracy Digital Thermometer User Reference Manual...
  • Page 2: Customer Service

    A rotatable probe mount allows for both horizontal and vertical orientation of the display module. A unique feature of the BetaProbe TI/TI+ is the trend indication arrows which the user can configure to show when the readings are considered to be stable enough to record the reading.
  • Page 3: Special Conditions For Safe Use

    Manufactured by Martel Electronics, Inc., under test: 3 Corporate Park Dr. Derry, NH, USA If the message changes to “OL ” the range limit is exceeded and the BetaProbe TI/TI+ must immediately be removed from the heat source to prevent damage.
  • Page 4: Normal Operation

    3. Operation Insertion Depth and Usage The BetaProbe TI/TI+ should be inserted to a minimum depth of 2.5" to obtain specified accuracy of measurement. The final 2 inches of the probe should always be exposed to prevent high temperatures from damaging the instrument's electronics or housing.
  • Page 5 Stabilty Display Stability Limit The stability display shows the general stability level relative to a user The stability limit setting is used to determine when the secondary configurable stability limit. There are 4 preset settings for stability display will indicate that the reading is stable. There are 4 preset represented in °C or °F (0.01, 0.1, 1.0, or 10.0).
  • Page 6 Note: The following Data Logging functions can only be ENTER Erase or cancel data erase. enabled with the optional BetaLOG-TI program which is available from Martel or your Martel distributor. Detailed information on the Logging functions is included with the BetaLOG software.
  • Page 7: Changing The Batteries

    Batteries must only be changed in an area known to be non-hazardous. To clean the BetaProbe TI/TI+ use a cloth with a mild cleaning solution. 6. Calibration The BetaProbe TI/TI+ can be calibrated in the field by one of three methods depending on the extent of the calibration needed.
  • Page 8: Specifications

    “Return Material Author ization” form can be obtained from our website www.martelcorp.com under customer Humidity Range: service. You can also call 1-800-821-0023 to have a form faxed. Martel 0 to 95% R.H. Non-condensing will not be responsible for damage as a result of poor return packaging.
  • Page 9 Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel Electronics 3 Corporate Park Drive Derry, NH 03038 0219876 Rev. D 11/11...