Instructions for Use AZOI_KITO_IFU_v1.33 Table of Contents About Kito ..........................5 Important Information ......................5 Safety ............................5 Intended Use ..........................5 Contraindication ........................5 Unpacking ..........................6 Package Contents ........................6 3.1.1 Device Unpacking ....................... 6 System Requirements for App installation ................6 Know your Device ........................
Page 4
AZOI_KITO_IFU_v1.33 About Azoi We, at Azoi, have been working on “monitoring technology” which would enable people to track their health. We intend to provide consumers an easy and accessible way to measure their critical vitals in a home environment. Azoi’s team includes experts in sensor technology, product design, application development and quality assurance to ensure that the delivered product conforms to the prescribed requirements.
(SpO2), respiration rate, heart rate and electrocardiographic activity (ECG). Kito uses a 1 Lead ECG mechanism to capture ECG; an NTC thermistor based sensor to measure skin temperature; and reflective SpO2 technology to measure blood oxygen saturation. Heart rate is calculated using the ECG values.
Instructions for Use AZOI_KITO_IFU_v1.33 3 Unpacking 3.1 Package Contents The package contains the Kito device and a Quick Start Guide along with the charger and a USB 2.0 cable as accessories. 3.1.1 Device Unpacking Open the box from the slide flap.
Connect the USB cable and turn ON the power supply. If you see a Blue light on the charger, Kito is charging. Indications The Switch (05) has a light which indicates the Kito battery level. Please refer to the following table to understand the indications: ...
4.3 Create an Account For the first time usage, you need to Sign-Up for creating your Kito account. To try Kito on the go, tap “Try It!” But before going any further, you will need to pair Kito with the app. Please see the pairing...
Note: Please turn Bluetooth ON from the “Settings” menu on your mobile device for pairing. You need to pair Kito with the Kito app only for the very first time. For subsequent usage, pairing is not required unless you use Kito with a different mobile device or change Kito.
You can choose to discard the current reading and take a reading again. 4.6 Remove Kito To remove Kito, use your thumbs to apply pressure on the corners of the Kito device and push it outwards. AZOI/FM/IFU/018 Rev 01...
5 Your Health Data 5.1 Know your Vitals Before proceeding further, let us quickly understand what the vitals measured by Kito mean: Heart Rate - Heart Rate is the number of times your heart beats in a minute and is an important indicator of cardiovascular fitness.
Rate for an adult in a relaxed state, ranges from 12 to 18 breaths per minute. Take a deep breath and read on. Note: A more detailed explanation regarding the vitals can be found in the Kito mobile app. Go to Settings ( and tap “About Vitals”...
ECG Filter, alter the measurement unit for the temperature vital and upgrade Firmware. Note: Kito and the mobile device must be sufficiently charged before upgrading the firmware. Once you initiate the firmware upgrade, you will not be able to access the application until the firmware installation is complete.
Page 15
The line graph which is a series of data points connected by lines, illustrates Fitbit™ data. For each Fitbit™ step count data, you can see the corresponding vital statistics taken from Kito overlaid on the graph.
Instructions for Use AZOI_KITO_IFU_v1.33 5.3.4 Skin Temperature 5.3.5 ECG ECG represents the electrical activity of the heart as measured by Kito’s electrodes. AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
6 Important Settings 6.1 Adding Multiple Users Kito supports multi-user access for up to four users per device. All users will have access only to their own individualized dashboards. Tap on the home screen to add a new user. To access an existing user’s dashboard, tap the username.
6.4 Sync Fitbit™ The Kito application overlays the information received from Fitbit™ on its own reading of vitals to indicate how your daily activity is affecting your vitals. It plots your Fitbit™ step count and Kito’s vital statistics on the same graph.
Page 20
Direct sunlight High temperature High humidity Wet or damp locations where water may get on the unit Dusty locations Near fires or open flames Locations with Strong vibrations Near strong electromagnetic fields AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
Unable to share the readings Check your internet connection. Unable to upgrade the firmware Kito is out of charge. Charge the device until the white indicator light stabilizes. Also confirm if your mobile device is connected to internet. Please note that in case you receive a call while upgrading firmware, the upgrade will not be completed.
Page 22
Instructions for Use AZOI_KITO_IFU_v1.33 Note: In the event that the above troubleshooting tips do not address your issue, please contact help@azoi.com. AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
1.2mV, given 1mV of input QRS amplitude Averaging Arithmetic mean over 30 seconds Response time 4sec for step increase from 80 BPM to 120 BPM 8sec for step decrease from 80 BPM to 40 BPM AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
Page 24
Contact based temperature sensing Minimum operating time 30 seconds Accuracy SpO2 +/- 4% rms Heart rate +/- 10% or +/- 5 BPM whichever is greater Respiration +/-3 BrPM Skin temperature +/- 0.3°C AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
Page 25
Wireless communication Bluetooth 4.0 - Bluetooth Low Energy Battery charging time from depletion 6 hours to 90% Charger Specifications Input USB 2.0 Output DC 5V+/-10%, @500mA, with 1KV DC-DC IEC60950 certified isolation AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
Page 26
2-Protection against vertically falling water drops when ENCLOSURE tilted up to 15° * Battery life varies by usage and settings. Symbols Description Refer to instruction manual/booklet Keep away from rain AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
11.1 Unpair You can remove the existing Kito pairing. Note: The pairing can be undone even if Kito is not in the vicinity. If you’ve lost Kito, you can unpair it and add a new Kito. Tap “Forget this Kito” and wait till the pairing is undone. To pair again, tap “Pair this Kito” and follow the on-screen pairing instructions.
Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
Consolidated version of 30 April 2015. 12.2 Electromagnetic Emissions Guidance and manufacturer’s declaration – electromagnetic emissions Kito is intended for use in the electromagnetic environment specified below. The user should ensure that it is used in such an environment. Emissions test...
IEC 61000-3-3 12.3 Electromagnetic Immunity Guidance and manufacturer’s declaration – electromagnetic immunity Kito is intended for use in the electromagnetic environment specified below. The user should ensure that it is used in such an environment. Electromagnetic Immunity test IEC 60601-1-2 test level Compliance level environment –guidance...
12.4 Guidelines and Manufacturer’s Declaration – Electromagnetic Immunity Guidelines and manufacturer’s declaration – electromagnetic immunity Kito is intended for use in the electromagnetic environment specified below. The user should ensure that it is used in such an electromagnetic environment. IMMUNITY test...
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may AZOI/FM/IFU/018 Rev 01 AZOI Proprietary and Confidential Page...
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and Kito Kito is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of Kito can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and Kito.
15 Declaration and Disclaimers The user is an intended operator. Kito may be interfered with by other equipment, even if that other equipment complies with the CISPR EMISSION requirements. Portable and mobile RF communications equipment can affect Kito.