Download Print this page

Lactate Pro 2 Operating Manual page 4

Blood lactate test meter

Advertisement

5.2 Adding a Delete Flag
Add a delete flag to inaccurate test results. The meter stores the test
results regardless of this flag. A delete flag indicates unnecessary data
when reviewing past results.
1. When the test result appears on the screen,
leave the test strip in the meter and press
button 2.
The delete flag "DEL" appears.
To remove the flag, press the button again.
2. To save the setting and turn off the meter, pull the test strip straight
out of the meter.
NOTE
• You cannot add/change the delete flag once the meter turns off.
• The meter turns off when you remove the test strip or after 3 minutes
from testing.
5.3 Reviewing Past Results
This system can store up to 330 test results. You can review the stored
test results in the order of the newest to the oldest.
1. Make sure the meter is off and test strip is not in the meter.
2. Hold down button 1 for 2 seconds.
A full display appears, followed by the present date and time. The
memory symbol appears at the top of the screen.
3. Press button 1 to select the measurer code you want to display.
The measurer code changes in the following order every time you
press the button:
ALL (all test results)
M01
M02
Total number of test results with the measurer code
The measurer code appears with the total number of test results for
that measurer code.
4. Press button 2.
The most recent test result for the selected measurer code appears.
Below the result, two sets of related information appear alternately as
shown below.
Test date
Measurer code
Data No.
Test time
NOTE
• When no test data is stored, "ALL" appears with "0". In this case, the
measurer code will not change even if you press button 1. Also, test
data will not appear even if you press button 2.
• To cancel call-up of test data, press button 2 for 2 seconds to turn off
the meter.
5. The test result appears from the newest to the oldest every time you
press button 1.
• • •
"End" appears after the oldest data in the memory. After "End", the
screen returns to the newest data.
To scroll through results, hold down button 1. The data No. does not
appear when scrolling through. "END" appears at the end of the scroll.
Every time you press button 2, the test result appears in reverse order
from the oldest to the newest.
6. To end this mode and turn off the meter, hold down button 2 for 2
seconds or leave the meter for 3 minutes.
5.4 Deleting Test Results
You can delete all the test results in the memory or results with specific
measurer code.
1. Make sure the meter is off and test strip is not in the meter.
2. Hold down button 1 for 2 seconds.
You will hear a beep when the meter turns on. A full display appears,
followed by the present date and time. The memory symbol appears
at the top of the screen.
3. Press button 1. Select the measurer code of
the test results you want to delete.
Select "ALL" to delete all the test results from
the memory.
4. Hold down both buttons 1 and 2 for 5 seconds.
Release the buttons after "dEL" flashes.
5. Hold down button 2 for 5 seconds.
The meter deletes the test results. Date and time appear again.
NOTE
• "dEL" stays on while the meter is deleting the test results.
This may take several minutes.
- 4 -
Chapter 6 Meter Setup
Set up the beeper volume and the date and time.
1. Make sure the meter is off and test strip is not in the meter.
2. Hold down both buttons 1 and 2 for at least 5 seconds.
A full display appears, followed by "SET".
 Beeper volume setting
3. Press button 1.
The present setting for the beeper volume
appears.
4. Press button 2.
The beeper volume changes in the following order every time you
press the button:
Level 3 (high)
OFF
Level 1 (low) Level 2 (medium)
The mute symbol appears when the beeper is set to OFF.
 Date/Time setting
5. Press button 1.
The "Year" setting screen appears.
6. Press button 2 to select the "Year".
To scroll through the years, hold down button 2.
7. Press button 1 to set the "Year" and go to
"Month" setting.
8. In the same way, set "Month", "Day", "Hour" and "Minute" in this
order.
When "Minute" is set, "END" appears.
• • •
9. To end the setup and turn off the meter, press button 1.
To return to the beeper volume setting, press button 2.
- 5 -