1、Introduction
Compact, rugged and easy to use. Just aim and push the
button, read current surface temperatures in less than a
second. Safely measure surface temperatures of hot,
hazardous or hard-to-reach objects without contact.
How it works
Infrared thermometer measures the surface temperature
of an object. The unit's optics sense emitted, reflected,
and transmitted energy which is collect and focused onto
a detector. The unit's electronics transmitted energy
which is display on the unit. For increased ease and
accuracy the laser pointer makes aiming even more
precise.
Cautions
Infrared thermometer should be protected for the
following:
--EMF(electro-magnetic fields) from arc welders,
induction heaters.
--Thermal shock(cause by large or abrupt ambient
temperature changes allow 1 hours for unit to stabilize
before use).
--Do not leave the unit on or near objects of high
temperature.
Warning
Do not point laser at eye or indirectly off reflective
surfaces.
1. When take measurement, point thermometer toward
the object to be measured and hold the yellow
trigger. The object under test should be large than
the spot size calculated by the field of view
diagram.
2. Distance & spot size: As the distance from the
object increase, the spot size of measuring area
becomes large.
3. Field of view: Make sure the target is larger than the
unit's spot size. The smaller the target the close
measure distance. When accuracy is critical, make
sure the target is at least twice as large as the spot
size.
4. Emissivity: Most organic materials and painted or
oxidized surfaces have an emissivity of 0.95.
Inaccurate readings will result from measuring
shiny or polished metal surfaces. To compensate,
cover the surface to be measured with masking tape
or flat black paint. Measure the tape or painted
surface when the tape or painted reach the same
temperature as the material underneath.
2、Quick start instruction
1. Press battery door clip, install battery correctly. Pull
the trigger, LCD display reading & battery icon.
Release the trigger and the reading will hold for 10
seconds
LCD display:
A Laser pointer turn on prompt
B Measure prompt
C Backlight prompt
D Data Hold
E
Temperature unit symbols
F Measure result
G
Low-voltage

Advertisement

loading

Summary of Contents for Delton 981A

  • Page 1 Non-contact infrared thermometer Warning 981A Do not point laser at eye or indirectly off reflective surfaces. Instruction manual 1. When take measurement, point thermometer toward the object to be measured and hold the yellow trigger. The object under test should be large than the spot size calculated by the field of view diagram.
  • Page 2 4、specifications 2. Locating a hot spot aim the thermometer outside the area of interest, then scan across with up and down Temperature -30℃ to 350℃(-22to 662 motions until you locate the hot spot.(please turn on range -30℃ to 450℃(-22 to 842 the laser to for accurate measuring) ±3% of rdg or ±3℃,-30℃...