it Key Points of the Prcedure ir
1.
While samples are being saved, the value indication disappears temporarily, and only the [M]
mark flashes. If the balance is affected by wind or other vibrations during this process, the
saving time may be prolonged.
2.
If [L " II r r ] appears briefly, it indicates one of the following states:
(1)
The weight of the reference sample is insufficient.
For the limit weight that can be saved (% limit weight), please refer to "13. Specifications," on
page 41.
(2)
While setting up the reference value in Step 2, the | Set | key has been pressed while the
samples were loaded on the balance.
$Klf [!_ - E r r ] appears, sampling has been interrupted and the sample value being processed
will not be saved.
3.
The minimum intervals between percentages in the unit switch from 1%, to 0.1%, to 0.01%,
depending on the reference weight from the sampling.
21