Denali M1 Thermoelectric Chiller
Manual
(from Solid State Cooling Systems built exclusively for Cole-Parmer)
167 Myers Corners Road, Wappingers Falls NY, 12569
(845)296-1300
email: info1@sscooling.com
www.sscooling.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Denali M1

  • Page 1 Denali M1 Thermoelectric Chiller Manual (from Solid State Cooling Systems built exclusively for Cole-Parmer) 167 Myers Corners Road, Wappingers Falls NY, 12569 (845)296-1300 email: info1@sscooling.com www.sscooling.com...
  • Page 2 52-13382-1 ENALI , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS APPINGERS ALLS : (845) 296-1303 F : (845) 296-1303 ELEPHONE SSCOOLING EVISION...
  • Page 3: Table Of Contents

    DVANCED PERATION 5.3 A _________________________________________________________________ 11 LARM IGNAL SECTION 6 ________________________________________________________________________ 12 ___________________________________________________ 12 YSTEM LARMS ROUBLESHOOTING SECTION 7 ________________________________________________________________________ 13 M1 RS-232 C __________________________________________________ 13 ENALI OMMUNICATION 7.1 C ___________________________________________________ 13 OMMUNICATION PECIFICATION 7.2 C _______________________________________________________ 14 OMMUNICATION ROTOCOL 7.3 RS232 C...
  • Page 4 CE Declaration of Conformity Solid State Cooling Systems 167 Myers Corners Road Wappingers Falls, NY 12590 declare under our sole responsibility that the Denali M1 meets the provisions of the directives: 2004/108/EC EMC Directive 2006/95/EC Low Voltage Directive EN 61326-1: 2006...
  • Page 5: Safety Precautions And Symbols

    52-13382-1 ENALI SAFETY PRECAUTIONS AND SYMBOLS _________________________ Read the MSDS for the coolant used and follow all safety precautions listed in the MSDS prior to removing coolant tubes or opening the fill cap as this could result in contact with the coolant inside.
  • Page 6: Section 1

    With fewer moving parts, the system is highly reliable and energy efficient. As the world’s smallest, air cooled recirculating chiller, the Denali M1 is ideal for laboratory use. It is the perfect bench-top solution to precisely control the...
  • Page 7: Section 2

    52-13382-1 ENALI ECTION _______________________________________________ PECIFICATIONS Operating (Set Point) Range: 0°C to 45 °C Ambient Temperature: 10°C to 40 °C (non-condensing) Stability / Repeatability: + 0.1 °C with constant load (even near ambient) Cooling Capacity: 180 watts @ 20°C in a 20°C ambient (See figure 1 for cooling curve) Heating Capacity: >240 watts @ 20°C in a 20°C ambient...
  • Page 8 52-13382-1 ENALI Figure 1 Denali M1 Capacity in 20°C Ambient Figure 2 Denali M1 Pump Performance @ 20°C Fluid Temperature Note: Pump to pump variation is +/- 10%. , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS...
  • Page 9: Section 3

    52-13382-1 ENALI ECTION ______________________________________________________ Figure 3A: Top View Keypad Display Screen Tank (Under Magnet Cover) 9-pin Dsub For RS-232 & Dry Contact Alarms Airflow Outlet Figure 3B: Back View ON/OFF Switch Coolant Return Coolant DC Power Supply Connector Airflow Inlet , 167 M , NY 12590 OLID...
  • Page 10: Electrical Connections

    NEMA L6-15 208 US Twist 16-23918-2 Alarms: The Denali M1 has one 250 VAC 1 amp dry contact relay to indicate a system alarm or temperature out of range. Connect to this dry contact on the 9-pin dsub connector as follows:...
  • Page 11: Air Considerations

    52-13382-1 ENALI 3.3 A ONSIDERATIONS Restricting airflow into or out of the Denali M1 will impair performance. Maintain at least 3” of clearance around the air inlet and outlet to ensure no restriction of airflow. 3.4 C OOLANT The coolant fill cap is located at the top of the unit. Twist and remove the plug then use the 250 ml bottle (shipped with the chiller) to fill reservoir prior to starting unit.
  • Page 12: Section 5

    5.2 A DVANCED PERATION The Denali M1 has two menus: the Status Menu and the Parameter Input Menu. The Status Menu shows the current temperature of coolant leaving the chiller (see Figure 4). The Status Menu also allows input of new coolant temperature set points when the cycling feature is off.
  • Page 13 52-13382-1 ENALI Figure 4: MENU STRUCTURE: OMENCLATURE UP or Increase Value Down or Decrease Value ↵ Press Enter Momentarily Press & Hold Enter Key 3 Sec SIMPLE OPERATION ADVANCED OPERATION press and hold enter key TATUS ARAMETER NPUT MENU TEMP: XX.X°C (current temp) PASSWORD XXXX ...
  • Page 14: Alarm Signal

    5.3 A LARM IGNAL The Denali M1 has one normally closed dry contact alarm for temperature out of range or system failure, located on pins 1 & 6 of the 9-pin dsub connector. A list of system failures causing the dry contact alarm to open can be found in Section 6.
  • Page 15: Section 6

    YSTEM LARMS ROUBLESHOOTING The Denali M1 has two system alarms that when triggered will show on the display. When an alarm is displayed the system will not attempt to heat or cool the coolant. TANK LOW: Liquid reservoir level is too low. Unless filling for WARNING the first time, check all outside plumbing lines for leaks.
  • Page 16: Section 7

    ______________________________ ENALI OMMUNICATION The Denali M1 comes with a modified (pin-out) RS232 communication port. The chiller uses this port to communicate a comprehensive set of control parameters with a Host PC - these parameters are outlined in Section 7.2. This port is a 9-pin female d- sub connector and is found on the front of the chiller (see Figure 3A).
  • Page 17: Communicationp

    52-13382-1 ENALI 7.2 C OMMUNICATION ROTOCOL Table 2: Command and Data Bytes Bit Position Description Bit =1 Bit = 0 Command Byte Bit 7 (MSB) Set Remote Control remote control local control (chiller) Bit 6 Remote On/Off chiller On chiller in standby Bit 5 Communication remote to chiller...
  • Page 18: Rs232 Communication

    52-13382-1 ENALI Notes 1) The Oasis can store 8 bytes of transmission. It is recommended that the transmissions be limited to 1 data exchange of one, two, or three bytes (depending upon the data being communicated), until the master has received the acknowledgement from the Oasis. 2) An acknowledgement of the transmission will be sent back to the master when the Oasis reads the data.
  • Page 19 52-13382-1 ENALI ECTION MSDS OOLANTS , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS APPINGERS ALLS : (845) 296-1303 F : (845) 296-1303 ELEPHONE SSCOOLING EVISION...
  • Page 20 52-13382-1 ENALI , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS APPINGERS ALLS : (845) 296-1303 F : (845) 296-1303 ELEPHONE SSCOOLING EVISION...
  • Page 21 52-13382-1 ENALI , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS APPINGERS ALLS : (845) 296-1303 F : (845) 296-1303 ELEPHONE SSCOOLING EVISION...
  • Page 22 52-13382-1 ENALI , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS APPINGERS ALLS : (845) 296-1303 F : (845) 296-1303 ELEPHONE SSCOOLING EVISION...
  • Page 23 52-13382-1 ENALI , 167 M , NY 12590 OLID TATE OOLING YSTEMS YERS ORNERS APPINGERS ALLS : (845) 296-1303 F : (845) 296-1303 ELEPHONE SSCOOLING EVISION...
  • Page 24 52-13382-1 ENALI MSDS for Ethylene Glycol ETHYLENE GLYCOL MSDS Number: E5125 --- Effective Date: 02/25/99 1. P RODUCT DENTIFICATION Synonyms: 1,2-Ethanediol; glycol; 1,2-Dihydroxyethane; Ethylene Alcohol; Ethylene Dihydrate CAS No.: 107-21-1 Molecular Weight: 62.07 Chemical Formula: CH2OHCH2OH Product Codes: J.T. Baker: 5387, 5845, 9140, 9298, 9300, 9346, 9349, 9356, L715 Mallinckrodt: 5001, 5037 2.
  • Page 25 52-13382-1 ENALI unconsciousness with convulsions. Death from respiratory arrest or cardiovascular collapse may follow. Lethal dose in humans: 100 ml (3-4 ounces). Skin Contact: Minor skin irritation and penetration may occur. Eye Contact: Splashes may cause irritation, pain, and eye damage. Chronic Exposure: Repeated small exposures by any route can cause severe kidney problems.
  • Page 26 52-13382-1 ENALI Special Information: In the event of a fire, wear full protective clothing and NIOSH-approved self-contained breathing apparatus with full-face piece operated in the pressure demand or other positive pressure mode. Toxic gases and vapors may be released if involved in a fire. 6.
  • Page 27 52-13382-1 ENALI Skin Protection: Wear protective gloves and clean body-covering clothing. Eye Protection: Use chemical safety goggles. Maintain eye wash fountain and quick-drench facilities in work area. 9. P HYSICAL AND HEMICAL ROPERTIES Appearance: Clear oily liquid. Odor: Odorless. Solubility: Miscible in water.
  • Page 28 52-13382-1 ENALI 11. T OXICOLOGICAL NFORMATION Toxicological Data: Oral rat LD50: 4700 mg/kg; skin rabbit LD50: 9530 mg/kg. Irritation - skin rabbit: 555mg(open), mild; eye rabbit: 500mg/24H, mild. Investigated as a tumorigen, mutagen, reproductive effector. Reproductive Toxicity: Has shown teratogenic effects in laboratory animals. --------\Cancer Lists\------------------------------------------------------ ---NTP Carcinogen--- Ingredient...
  • Page 29 52-13382-1 ENALI --------\Federal, State & International Regulations - Part 1\---------------- -SARA 302- ------SARA 313------ Ingredient RQ TPQ List Chemical Catg. ----------------------------------------- --- ----- ---- -------------- Ethylene Glycol (107-21-1) No No --------\Federal, State & International Regulations - Part 2\---------------- -RCRA- -TSCA- Ingredient CERCLA 261.33...
  • Page 30 52-13382-1 ENALI Disclaimer: Mallinckrodt Baker, Inc. provides the information contained herein in good faith but makes no representation as to its comprehensiveness or accuracy. This document is intended only as a guide to the appropriate precautionary handling of the material by a properly trained person using this product.
  • Page 31: Msds For Coolants Warranty Policy

    ARRANTY OLICY The Denali M1 Chiller is covered under a one-year parts and labor warranty from the date of shipment, assuming proper use and maintenance of the unit. All warranty work shall be performed at Solid State Cooling Systems’ facility, currently located in Wappingers Falls, NY, USA and requires pre-authorization by SSCS.