6.3

Exposing the Storage phosphor screen

Step
1
2
3
4
Getting Started with Typhoon FLA 7000 28-9607-64 AD
Action
Erase the Storage phosphor screen completely.
Wrap the radio isotope sample with plastic film. Make sure not to wrinkle
the wrapping film.
Note:
When using a tritium storage phosphor screen to detect tritium, place the
sample directly on the storage phosphor screen. A tritium storage phosphor
screen can be used only once.
Open the top cover of the cassette. Unlock the cassette by sliding the buttons
on both sides up.
Place the sample on the cassette. Make sure that the sample surface faces
up.
Note:
Keep the radio imaging sample away from the edge of the Storage phosphor
screen. Otherwise, the recorded data may become corrupted.
6 Exposing Storage phosphor screens
6.3 Exposing the Storage phosphor screen