Finishing the Adjustment

To Read the Calibration Count

N O TE
Agilent U1251B/U1252B User's and Service Guide
1 Remove all the shorting plugs and connectors from the
instrument.
2 Record the new Calibration Count.
3 Press
and
Adjustment Mode. Power off and on again. The instrument
will then be secured.
You can query the instrument to determine how many
calibrations have been performed.
Your instrument was calibrated before it left the factory.
When you receive your instrument, read the count to
determine its initial
The count value increases by one for each calibration point,
and a complete calibration will increase the value by many
counts. The calibration count increments up to a maximum
of 65535 after which it is set at 0. The calibration count
can be read from the front panel after the instrument has
been unsecured. Use the following procedure to read the
calibration count from the front panel.
1 Press
Adjustment Mode. The primary display shows
the calibration count.
2 Take note of the count.
3 Press
again to exit the calibration count mode.
Performance Tests and Calibration
simultaneously to exit the
value.