Summary of Contents for Omega HHM14

  • Page 1 OMEGA...
  • Page 2 European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Safety Information

    SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: 1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly. 2.
  • Page 4: Specifications

    SPECIFICATIONS Display: 4½ digit liquid crystal display (LCD) with a maximum reading of 19999. Polarity: Automatic, positive implied, negative polarity indication. Overrange: (1) or (-1) is displayed. Zero: Automatic. Low battery indication: the " " is displayed when the battery voltage drops below the operating level.
  • Page 5 DC VOLTS Ranges: 200mV,2V,20V,200V,1000V Resolution: 10µV Accuracy: ±(0.05%rdg + 4dgts) Input impedance: 10MW Overload protection: 1000VDC or 750VAC rms AC VOLTS (True RMS) Ranges: 200mV,2V,20V,200V,750V Resolution: 10µV Accuracy: Range Input 45Hz-1KHz 1KHz-20KHz 200mV 20mV-200mV 200mV-2V ±(1.0%rdg+10dgts) ±(1.5%rdg+10dgts) 2V-20V 200V 20V-200V ±(1.5%rdg+10dgts) 750V 200V-750V...
  • Page 6 DC CURRENT Ranges: 200µA,2mA,20mA,200mA,10A Accuracy: ±(0.5%rdg + 10dgts) on 200µA to 200mA ranges ±(2.0%rdg + 10dgts) on 10A range Input protection: 0.5A / 250V fast blow fuse 10A / 600V fast blow ceramic fuse AC CURRENT (True RMS 50Hz - 500Hz) Ranges: 200µA,2mA,20mA,200mA,10A Accuracy: ±(1.0%rdg + 10dgts) on 200µA to 200mA ranges ±(2.5%rdg + 10dgts) on 10A range...
  • Page 7 CONTINUITY Audible indication: less than 40W ±20W Overload protection: 500VDC or AC rms DIODE TEST Test current: 1.0mA ± 0.6mA Accuracy: ±(2.0%rdg + 3dgts) Open circuit volts: 3.3Vdc typical Overload protection: 500VDC or AC rms TRANSISTOR hFE Ranges: 0 - 1000 Base current: 10µAdc approx.
  • Page 8: Operation

    OPERATION Before taking any measurements, read the Safety Information Section. Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements.
  • Page 9 5. Use caution when measuring 10 amps on 10A range for 60s, please waiting for 10 minutes for next measurement of 10 amps for safety reason. Resistance and Continuity Measurements 1. Set the Function/Range switch to the desired resistance range or continuity position.
  • Page 10 Diode Tests 1. Connect the red test lead to the "VW " jack and the black test lead to the "COM" jack. 2. Set the Function/Range switch to the " " position. 3. Turn off power to the circuit under test. 4.
  • Page 11: Maintenance

    MAINTENANCE WARNING Remove test leads before changing battery or fuse or performing any servicing. Battery Replacement Power is supplied by a 9 volt "transistor" battery. (NEDA 1604 IEC 6F22). The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery case.
  • Page 12 CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any...
  • Page 13: Data Acquisition

    Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE DATA ACQUISITION PRESSURE/STRAIN AND FORCE HEATERS FLOW/LEVEL ENVIRONMENTAL MONITORING AND CONTROL pH/CONDUCTIVITY M-2886/0799...