1.4 (Battery Type Selection)
This setting determines whether Alkaline, NiCad (or NiMH) batteries are used for battery
operation.
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are: 0, 1, or 2.
0
Alkaline batteries - battery charging is DISABLED
1
NiCad or NiMH batteries - battery charging is ENABLED
2
Battery charging is ENABLED and FORCED ON. This selection
forces battery charging for NiCad or NiMH batteries that are
discharged.
CAUTION! Selecting 1 or 2, enables battery charging. DO NOT select 1 or 2,
when using Alkaline batteries.
+"55%?a(Sleep Mode)
The Sleep Mode feature also conserves battery power when the indicator remains unused for
a period of approximately 1 to 9 minutes. When enabled, the load cell excitation will be
reduced and the display will show +"%. The Sleep feature requires the indicator to remain
at the center of zero to activate, unlike Automatic Shutoff feature which only requires no
motion. Weight placed on scale will activate indicator and return it to weight mode.
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are 0 through 9 with 0 disabling Sleep Mode
+ ?a(Automatic Shutoff)
The Automatic Shutoff feature will turn the indicator off after a period of approximately 1 to 9
minutes of inactivity to prolong battery life. You must press the ON/OFF key to turn indicator
back on.
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are 0 through 9 with 0 disabling Automatic Shutoff.
Setup and Calibration Is Completed
The setup and calibration process has been completed. The indicator will reset and then turn
off. Remove power from the indicator and re-assemble for use.
8555-M260-O1 750 Indicator
SETUP AND CALIBRATION, Cont.
Page 15