SECTION 2—COMPONENT REPLACEMENT
1. Remove label from the top of the XPO
 (FIGURE 2.2). 
2
NOTE: This label adheres to the front and rear sections of the XPO
. This label MUST be removed 
2
to separate the two sections.
Label
XPO
2
FIGURE 2.2 Remove/Install label
2. Position the XPO
 so that the back panel label is facing up (Detail "A" in FIGURE 2.3).
2
®
3. Remove the four mounting screws from the back of the XPO
 using a T15 Torx
 
2
screwdriver (Detail "A" in FIGURE 2.3).
DETAIL "B"
DETAIL "A"
Mounting Screws
Ribbon Cable
Front Cover
Inlet Filter
Back Panel Label
Tubing
Mounting Screws
Front Cover
DETAIL "C"
Ribbon Cable
PC Board
FIGURE 2.3 Removing the Front Cover
10
XPO2™
Part No 1150748