Allpro ScanX-12 DVM Instruction Manual page 21

Digital imaging system in-line erase
Table of Contents

Advertisement

INTRAORAL IMAGING PROCEDURES
Scanning and Erasing Plates (continued)
Up to four PSPs can be processed simultaneously. One PSP can be inserted into each
Note:
Plate Guide at a time as long as the corresponding track indicator light is illuminated
green. The next PSP may be fed into a Plate Guide slot only after the corresponding track
indicator LED changes from yellow to green.
6. Observe that a red glow eminates from the scanner exit slot.
7. Repeat steps 2 through 6 to process additional PSPs as necessary. Another
PSP may be fed into any Plate Guide slot as long as the track indicator
light illuminates green.
8. Observe that each scanned and erased PSP drops into the receiving tray
at the bottom of the scanner. Since the ScanX default operation is with
the erase mode enabled (blue LED indicator above the ERASER switch is
steadily illuminated), each PSP is erased and ready for reuse.
9. Observe that all transport indicators illuminate green and the red glow
from the exit slot extinguish after the last PSP drops to the tray.
CAUTION:
10. Retrieve the processed (scanned and erased) PSPs for reuse or storage.
Make sure not to scratch the sensitive surface or nick the edges when
removing from the scanner outlet.
11. View and save each scanned image using the user-supplied authorized
Imaging Software.
IMPORTANT:
Scanning Plates without Erasing
The ScanX can be operated with the in-line eraser feature turned off. When the
eraser mode is disabled, the ScanX scans the same as when the eraser is enabled
except that the PSPs are not erased after scanning. Scan an intraoral PSP without
erasing the image by performing the following procedures.
1. Activate the scanner by performing the pro-
cedures on previous page.
Upon activation, the ScanX defaults with the
Note:
eraser mode enabled. This must be disabled
prior to scanning to prevent erasing of the
scanned PSP.
2. Disable the eraser mode of operation by press-
ing the membrane ERASER switch located on the
Keypad Control Panel. See Figure 10.
3. Verify that the blue LED indicator above the
ERASER switch blinks blue to indicate that
the Erase function is OFF. PSPs will not be
erased after scanning.
4. Insert the exposed PSPs to be scanned into the ScanX Plate Guides by
performing the Scanning and Erasing Plates procedures provided on the
previous page.
5. The scanned PSPs still contain latent images that require erasure. Make
sure to erase each PSP prior to reuse for new images.
Always use exercise care in handling the PSPs so as not to scratch the
sensitive surface or nick the edges.
PSPs will not be erased after scanning when operating the ScanX
with the eraser disabled. PSPs must always be erased prior to expo-
sure to X-rays for new images.
ScanX Keypad Control Panel
21
GREEN LED
Indicator
READY
READY
ALLPRO
ALLPRO
IMAGING
IMAGING
READY
Switch
Figure 10.
BLUE LED
Indicator
FLASHING
FLASHING
OFF
OFF
ERASER
ERASER
ERASER
Switch

Advertisement

Table of Contents
loading

This manual is also suitable for:

Scanx dvm

Table of Contents