Install the Reliant Scale
NOTE: For this procedure, refer to FIGURE 2.
1. Position the load cell assembly of the Reliant Scale into the boom mounting bracket. Refer to Detail 
ʺAʺ in FIGURE 2.
NOTE: For 9805 lift use 1/8‐inch thick nylon washer. For 450 lbs/600 lbs lifts use 1/4‐inch thick nylon washer. 
2. Secure the Reliant Scale to the boom mounting bracket with a shoulder bolt, one small steel washer, 
two nylon washers and a locknut. Securely tighten. Refer to Detail ʺAʺ in FIGURE 2 for washer 
orientation.
NOTE: Ensure the shaft of the shoulder bolt passes through both sides of the boom mounting bracket.
3. Insert swivel bar pin with two nylon washers through the swivel bar. Refer to Detail ʺBʺ in 
FIGURE 2.
4. Align the mounting holes in the swivel bar pin with the mounting holes in the load cell assembly. 
Refer to Detail ʺBʺ in FIGURE 2.
5. Secure swivel bar pin to the load cell assembly with the provided mounting screw and locknut. 
Securely tighten. Refer to Detail ʺBʺ in FIGURE 2.
6. Weigh the patient. Refer to Operating the Scale on page 4.
DETAIL "A" - ATTACHING
Locknut
Boom
Mounting
Bracket
Nylon Washer
Reliant
Scale
Part No 1114832
THE SCALE
Boom
Shoulder
Nylon
Washer
Load Cell
Assembly
FIGURE 2 Install the Reliant Scale
DETAIL "B" - ATTACHING THE
Reliant Scale
Mounting
Screw
Bolt
Nylon
Washers
3
HANGER BAR
Locknut
Swivel Bar
Swivel Bar Pin