Invacare TRACER EX Operator And Maintenance Manual page 27

Invacare wheelchair owner's operator and maintenance manual
Table of Contents

Advertisement

OPTIONS
Adjusting Front Anti-tippers
1. Measure from the floor to the bottom of the wheels
of the triangular assembly.
2. Perform one (1) of the following:
TRIANGULAR ASSEMBLY 3/4-INCH FROM
FLOOR - Repeat STEP 1 for the opposite triangular
assembly.
TRIANGULAR ASSEMBLY NOT 3/4-INCH FROM
FLOOR - Proceed to STEP 3.
3. Remove the machine screw, washers and locknut
securing the triangular assembly to the bottom of
the anti-tipper arm.
4. Position the triangular assembly mounting hole to
align with one (1) of the adjustment holes in the anti-
tipper arm.
5. Measure from the floor to the bottom of the wheels
of the triangular assembly.
INSTALLING
Wheelchair Frame
Bracket Clamps
Locknuts
Triangular Assembly
NOTE: Right side Front Anti-Tipper pictured.
FIGURE 5 - INSTALLING/ADJUSTING THE FRONT ANTI-TIPPERS
6. Repeat STEPS 4-5 until the bottom of the wheels of
the triangular assembly measure approximately
3/4-inch from the floor.
7. Secure the triangular assembly to the bottom of the
anti-tipper arm with the machine screw, washers and
locknut.
DO NOT overtighten the locknut securing the tri-
angular assembly to the bottom of the anti-tip-
per arm. Otherwise, anti-tipper arm will not swing
freely and the wheelchair may tip over, resulting
in injury or damage.
8. Tighten the locknut until the triangular assembly does
not rattle, but swings freely at the end of the anti-
tipper arm.
9. Repeat STEPS 1-2 for the opposite triangular as-
sembly.
Lug Bolts
Front Caster
Adjustment
Attaches Here
Holes
(not shown for
clarity)
Locknut
Anti-Tipper
Arm
27
WARNING
ADJUSTING
Anti-Tipper Arm
Washer
Triangular
Assembly
é
3/4-inch
ê
FLOOR
PROCEDURE 8
O
P
T
I
O
N
S
Machine
Screw
Mounting
Holes
Wheel

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents