To Check/Replace Relay; Overload Protector; To Check/Replace Overload Protector; Run Capacitor - Electrolux Wine Cooler & Beverage Center Service Manual

Wine cooler & beverage center with addendum added for refrigerator drawers
Hide thumbs Also See for Wine Cooler & Beverage Center:
Table of Contents

Advertisement

C - Electrical Components / Operation

To Check/Replace Relay

1.
Disconnect electrical supply to refrigerator.
2.
Remove bale wire holding relay to
compressor.
3.
Remove relay assembly from compressor.
(See Figure C2.)
4.
Use small, flat-bladed screwdriver to
disconnect leads to relay assembly.
5.
Use flat headed screwdriver to gently pry
capacitor from relay assembly.
6.
Use ohmmeter to check resistance between
terminals 5 and 6. Resistance should be 3
to 12 ohms at normal room temperature.
Shorted relay will read 0 ohms. Open relay
will read very high or infinite resistance.
7.
If ohm readings are out of range, replace
relay.
8.
Reverse this procedure to re-assemble.
NOTE
When replacing leads to the PTC relay, ensure
locking tabs snap into terminal.

Overload Protector

The overload protector is completely thermally
operated. It will open from excessive heat or
current. Unlike prior overloads, the internal bi-
metal is not self-heating, and is not a part of the
electrical circuit. The overload has a small built-in
coil heater that is in series with the compressor
start and run windings (See Figure C3).
To Check/Replace The Overload
Protector
1.
Disconnect electrical supply to refrigerator.
2.
Remove bale wire holding relay to
compressor.
3.
Remove relay assembly from compressor.
4.
Use flat headed screwdriver to gently pry
capacitor from relay assembly.
5.
Use small, flat-bladed screwdriver to
disconnect leads to relay assembly. (Note:
On some models you will have to remove
bale wire and cover, to gain access to relay
and overload protector.)
6.
Use ohmmeter to check resistance between
tab terminal and female pin terminal.
Overload protector should have less
than 1 ohm of resistance at normal room
temperature.
7.
If ohm readings are out of range, install new
Starter/Overload Assembly. See Figure 2.
8.
Reverse this procedure to re-assemble.
NOTE
When replacing leads to the PTC relay, ensure
locking tabs snap into terminal.

Run Capacitor

The run capacitor has permanently attached
terminals which are connected to relay terminals 2
and 3.
NOTE
Some models are not equipped with a run
capacitor.
To Check/Replace The Run Capacitor
1.
Disconnect electrical supply to refrigerator.
2.
Remove bale wire holding relay to
compressor.
2.
Use small, flat-bladed screwdriver to
disconnect leads to relay assembly.
3.
Use flat-bladed screwdriver to gently
pry capacitor from relay assembly.
4.
Discharge capacitor by shorting across
terminals with 500K (1 watt) resistor for one
minute.
Use ohmmeter set on the "Ohms times
5.
1000" scale (if available), to check resistance
across capacitor wire terminals.
• The needle should jump towards zero ohms
and quickly move back to infinity.
• If the needle does not move, the capacitor is
open.
• If the needle reads a constant value at or
near zero ohms, the capacitor is shorted out.
• If the needle jumps toward zero and then
moves back to constant high resistance (not
infinity), the capacitor has a high resistance
leak.
6.
If ohm readings are out of range, replace
capacitor.
7.
Reverse procedures to re-assemble.
13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents