Accessories Required - HP 3325A Calibration Manual

Synthesizer/function generator
Table of Contents

Advertisement

5. Accessories Required.

usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
50
feedthrough termination, Hewlett-Packard, Model 11048C.
Common name
AUDIO ANALYZER
FREQUENCY COUNTER
MEASURING RECEIVER
MULTIMETER
OSCILLOSCOPE
SPECTRUM ANALYZER
SYNTHESIZER/LEVEL
GENERATOR
TIME FREQUENCY
WORKSTATION
RESISTANCE STANDARD
The accessories required for this calibration are common
Table 2. Minimum Specifications of Equipment Required
Minimum use specifications
Output: 4.5 V, 10 kHz
AM distortion:
-60 dB at 50 Hz
Range: 99,9950 s to 60,000,300 Hz
Accuracy: 1.25 x 10
Capability: Time interval A to B
Frequency: 100 kHz to 10 MHz
Volts: .683 to .732 V rms
Accuracy:
AM: 0 to 80%
Range: -20.225 to +20.225 V dc
Accuracy:
Range: 0.3416 to 20.4 V ac
100 Hz to 100 kHz
Accuracy:
Frequency: 1 kHz to 10 MHz
Amplitude: 1.8 to 2.2 V p-p
Accuracy:
2.5%
Capabilities
Duty cycle: 49.7 to 50.3%
Overshoot: 5%
Rise time and fall time: <20 ns
Frequency: 50 kHz to 20 MHz
Input: -20 dBm and 40 V
Display capability: <-65 dBc
Frequency:
Amplitude: 13 dBm
Reference output frequency: 10 MHz
Accuracy:
Range: 470
-30 dB at 10 kHz Distortion:
-8
0.85%
0.05%
0.3%
0.1 MHz
5 x 10
-8
TB 9-6625-2049-35
Manufacturer and model
(part number)
Boonton, Model 1120-S/10
(MIS-35954/2)
Hewlett-Packard, Model 5345A
(MIS-28754/1 Type 1)
Consisting
of
:
Measuring
receiver
Hewlett-Packard,
Model
8902A
and
Sensor
module Hewlett-Packard, Model
11722A
Hewlett-Packard, Model
3458A
Tektronix, Type 2430A
(OS-291/G)
(AN/USM-489A(V)1)
Hewlett-Packard, Model
3335AOPT 001-K06
(MIS-35938)
Autek Systems Corporation,
Model 620 (MIS-38946)
Biddle Gray, Model 71-631
(7910328)