5.3 Troubleshooting

PLP-701 / PLP-702 / PLP-701R / PLP-702R
Rotor
Does not rotate
NOTE:
*
• When the operation is continued, set the laser beam on the reference
mark and perform it.
Manual mode display lamp:
Display
Display lamps are
not turned on.
Lamp
turns on.
1
Lamp
& lamp
2
1
blink mutually
MANUAL
MANUAL
BATTERY
BATTERY
MODE
MODE
CHECK
CHECK
Content
Battery capacity
1
is low.
Incorrect battery
2
insertion.
Instrument's tilt
1
exceeds auto leveling
area. (± 5.7°)
Instrument is tilt
2
largely by any cause.
After power supply
3
turns on, unstable
state of auto
self-leveling function
continues
approx.3 min.
NOTE*
Defect of instrument
itself.
2
OFF
OFF
LEVEL
LEVEL
Off level display lamp:
POWER
POWER
ON
ON
OFF
OFF
Checking point
Replace with fresh
1
batteries or charge
battery
Check the polarity
2
of the batteries.
Turn off the power
1
supply and reset the
instrument level
(± 5.7°) and turn on
again.
Turn off the power
2
supply and turn on
again.
Turn off the power
3
supply and turn on
again.
Turn off the power
supply and turn on
again. If the lamps blink
farther, instrument need
to be repaired.
1
37