Agilent Technologies 11970 Series User Manual page 52

General-purpose harmonic mixers 11970 series
Hide thumbs Also See for 11970 Series:
Table of Contents

Advertisement

Performance Tests
Performance Test Procedures
Table 3-2. Recommended Test Equipment for 11970A
Instrument
Spectrum Analyzer
Synthesized Sweeper
Amplifier
Amplifier
Power Meter
Power Sensor
Power Sensor
1
Directional Coupler
Isolator
Frequency Tripler
Adapter
Adapter
Adapter (2 required)
Cables (3 required)
Cable
Cable
1. Calibration data for the coupling ration between the output arm and the auxil-
iary arm is necessary for accurate measurements.
2. Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA
52
Critical Specifications
LO and IF ranges compatible with mixer
Frequency: 8 to 13.5 GHz
Output Level: >0 dBm
Output Level: >18 dBm leveled
Frequency Range: 3 to 5 GHz
Frequency Range: 8 to 13.5 GHz
Output Level >15 dBm
Compatible with Power Sensor
SWR: <2.0
SWR: <1.3 @ 6 GHz
Coupling: 20 dB
Directivity: >40 dB
Primary Arm SWR: <1.05
Auxiliary Arm SWR: <1.2
Insertion Loss: <1.5 dB
Isolation: >20 dB
SWR: <1.2
Input Power: >10 dBm minimum
Conversion Loss: <15 dB
SMA female to SMA female
SMA female to Type N female
Type N male to SMA female
Connectors: SMA male
Connectors: SMA male
Loss: <1.0 dB @ 20 GHz
Connectors: BNC male
Recommended
Model
8566B
8340A
11975A
8349A
436A
R8486A
8485A
R752D
P/N 0960-0082
2
Spacekom Microwave
TKa-1
P/N 1250-1158
P/N 1250-1404
P/N 1250-1250
P/N 5061-5458
P/N 8120-4396
11086A
Chapter 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

11970k11970a11970q11970u11970v11970w

Table of Contents