Mounting The Integral Modem (Ecat-Pt) - Honeywell Mercury User Manual

Table of Contents

Advertisement

Page 12
www.honeywell.com
12
Mounting the Integral Modem
(ECAT-PT)
Before Installation
1.
Start the AT-Link Software package and under Instrument
on the main menu line, select Shutdown. All data is stored in
memory and the instrument is not obtaining data.
2.
When Shutdown is complete, reach inside the instrument and
unplug the main battery plug.
3.
Remove any battery or power supply mounted in the right-
hand position.
4.
Remove the (2) hex nuts holding the door latch to the case. Do
not remove the latch or chassis bumper assembly.
After Installation
Mercury Modem with
Main Batteries not
shown for clarity
Figure 3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents