Configuring the CodeMaster 100 and 12-Lead ECG Option

Configuring the CodeMaster 100 and 12-Lead ECG Option
2
This section provides instructions for configuring the CodeMaster 100 and 12-Lead
ECG option.
To configure general items such as the date and time, as well as all defibrillating and
monitoring settings, use the Setup Menus accessible through the front panel keys
and front display. Table 2-1 and Table 2-2 list the selections available on the Setup
Menus.
To configure 12-Lead ECG and transmission settings, use the Configuration Menu
accessible from the top keypad and upper display. Table 2-3 lists the selections
available on the Configuration Menu.
A small internal battery provides back up power to store the configuration
information. Under normal conditions, the backup battery may last for five years.
Extreme conditions (such as prolonged high temperatures) may reduce the life of the
backup battery.
After running the "Delivered Energy and Shock Button Functional Test" on page
13-3, you may see a CHECK SETUP message on the screen. This message appears
when the backup battery has approximately 90 days of life remaining.
When you see the CHECK SETUP message on the screen, call service to replace the
CAUTION
backup battery.
If you do not replace the backup battery, the SETUP LOST message appears on the
screen after the internal battery depletes, and you may have to reconfigure the
instrument.