Download Print this page

Specifications; Ordering Information; General Specifications - Fluke 726 Technical Data

Precision multifunction process calibrator
Hide thumbs Also See for 726:

Advertisement

726 Specifications

Function
Measure or Source
Voltage
0 to 30 V (measure)
mA
mV (TC terminals)
-10.00 mV to +75.00 mV
Resistance
Frequency
Loop Supply
Thermocouples
J, K, T, E, L, N, U, C, BP, XK
Thermocouples
RTDs
Cu 10 Ni 120 (672)
Pt 100 (3926) Cu 10
Simultaneous Function Capability
24.000 mA DC
24.000 mA DC with 24 V loop supply
100.00 mV DC
30.000 V DC measure
20.000 V DC measure, 20.000 VDC Source
5 to 4000 Ohms
Thermocouple J, K, T, E, R, S, B, M, L, U, N, C, BP, XK
RTD Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200,
500, 1000 (385), Cu 10
Pressure (requires Fluke 700PXX Modules)
Frequency; Squarewave, 1 CPM to10 kHz;
fixed amplitude 5 V p-p
M = Measure, S = Source/Simulate

General Specifications

Storage temperature: -20 °C to 71 °C
Operating temperature: -10 °C to 55 °C
Relative humidity: 90 % (10 °C to 30 °C);
75 % (30 °C to 40 °C); 45 % (40 °C to 50
°C); 35 % (50 °C to 55 °C)
Shock: 1 meter drop test
Safety: CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and
EN55022:1994 Class B
Size (HxWxD): 200 mm x 96 mm x 47 mm
(7.9 in x 3.8 in x 1.9 in)
Weight: 650 g (23 oz)
Battery: Four AA alkaline batteries
Battery life: 25 hours typical
Warranty: Three-years
2 Fluke Corporation
726 Precision Multifunction Process Calibrator
Range or Type
Resolution
0 to 100 mV
0 to 20 V (source)
0 to 24
0.001 mA
5 Ω to 4000 Ω
≤ 0.01 Ω to 0.1 Ω
2.0 to 1,000 CPM
1 to 1100 Hz
1.0 to 10.0 kHz
10.0 to 15.0 kHz
24 V dc
0.1 ºC, 0.1 ºF
B, R, S
1 ºC, 1 ºF
0.1 °C, 0.1 °F
Pt 100, 200, 500,
1000 (385)
Pt 100 (3916),
0.01 ºC, 0.01 °F
Channel A

Ordering Information

Fluke-726
Included
TL75 Test Leads, AC72 Test Clips, one pair
of stackable test leads, product overview
manual (print) and user's manual (CD-ROM)
in14 languages.
Accuracy
0.01 mV
0.01 % Rdg + 2 LSD
0.01 V
0.01 V
0.01 % Rdg + 2 LSD
.01 mV
0.01 % of range + 1 LSD
0.015 %
0.1 CPM
± 0.05 %
1 Hz
± 0.05 %
0.1 kHz
± 0.25 %
0.1 kHz
± 0.5 %
N/A
10 %
to 0.2 ºC
to 1.2 ºC
to 1.8 °C
to 0.15 ºC
Channel B
M
M or S
M
M or S
M
M or S
M or S
M or S
M or S
M
M used as S
M or S
Precision Multifunction
Process Calibrator
Notes
Max load, 1 mA
Max load, 1000 Ω
For frequency source;
1 V to 20 V p-p
squarewave, -0.1 V offset
Fluke. Keeping your world
up and running.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2005 Fluke Corporation. All rights reserved.
Printed in U.S.A. 12/2005 2527590 D-EN-N Rev C

Advertisement

loading