Set Auto Off; Show Battery Voltage; Display Actual Temperature; Set Damping - Fluke 717-10000G User Manual

Pressure gauge
Hide thumbs Also See for 717-10000G:
Table of Contents

Advertisement

Set Auto Off

Auto Off can be set in 1-minute increments from 1 to 30
minutes or you can turn off the function for continuous
Product operation. The Product is configured for 30
minutes. Push  and  to set the necessary interval.
The "off" position is at the low end of the selections, less
than 1 minute.

Show Battery Voltage

Actual battery voltage and a percent-of-life bargraph
show the battery charge. No adjustments are made in this
parameter.

Display Actual Temperature

The Product is temperature compensated This parameter
shows the temperature measured by the internal sensor.
Push  or  to show degrees F or C.

Set Damping

Selections are "on"  and "off" . Damping smooths
readings from pulsating pressure sources.
Shop for Fluke products online at:

Set Sample Rate

This function finds how often pressure is sampled and the
display is updated. Selections are 0.5, 1, 3, and 10
samples/second. Note that 10/second gives the fastest
response time.

Set Tare

Use this function to set a constant offset value which is
then subtracted from the measured pressure. For
example, if a tare is set at 30 psi, and the measured
pressure is 37 psi, 7 psi is shown.
A pressure of 27 psi is shown as -3 psi.
Push  and  to set the tare value. The value, is based
on the engineering units and resolution selected for
display. Tare value can be set to the maximum range of
the gauge.
For safety, the bar graph always shows the actual
pressure based on the full range of the gauge regardless
of the tare position. This is done to make sure that even
with a "0" reading pressure is being applied to the gauge.