Agilent Technologies U1231A User Manual page 125

Handheld digital multimeter
Hide thumbs Also See for U1231A:
Table of Contents

Advertisement

Table 5-1 DC specifications (continued)
Function
Range
60 μA
[1]
600 μA
[1]
Current
[2][4]
6 A
[2][3]
10 A
Notes for DC current specifications:
1 Overload protection for 60 μA to 600 μA range: 600 Vrms for short circuits with <0.3 A current.
2 Overload protection for 6 A to 10 A range: 11 A/1000 V; 10 × 38 mm fast-acting fuse.
3 Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 A to 20 A
for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the
measured time before proceeding with low current measurements.
4 DC current range of 0.6 mA to 1 mA is not measureable on the U1232A and U1233A models.
U1231A/U1232A/U1233A User's Guide
Accuracy
Resolution
U1231A
0.01 μA
-
0.1 μA
-
0.001 A
-
0.01 A
-
Characteristics and Specifications
U1232A
U1233A
1.0% + 2
1.0% + 2
1.0% + 2
1.0% + 2
1.0% + 3
1.0% + 3
1.0% + 3
1.0% + 3
Electrical Specifications
Test
Burden
current
voltage
(where applicable)
-
<2.5 V
-
<2.5 V
-
<0.2 V
-
<0.4 V
5
Input
impedance
-
-
-
-
107