Procedure; Transmitted Pdh Jitter Amplitude Accuracy On Range 1 - HP 37717C Calibration Manual

Communications performance analyzer
Hide thumbs Also See for 37717C:
Table of Contents

Advertisement

C AU T I O N
Performance Tests
Transmitted Jitter/Wander Amplitude Accuracy (Option A3K [A3Q])
3. Make the following key sequence on
the HP 37717C to obtain the special
display.
MODULE DEBUG
Press
;
;
OTHER
MORE
;
;
MORE
MORE
Press
until
MORE
MODULE DEBUG
appears in the softkey menu.
Press
MODULE DEBUG
MODULE PDH Module.
Set the VCO Control Mode to FIXED.
The sequence above must be performed each time the power is cycled on the HP
37717C, as all
MODULE DEBUG
is cycled.

Procedure

Transmitted PDH Jitter Amplitude Accuracy on Range 1

1. Select
TRANSMIT
37717C display as shown opposite.
3-136
;
;
;
OTHER
.
and select
parameters adopt DEFAULT values when power
and setup the HP