SOURce:PHASe<x>:VOLTage:FLICker:DUTY(?) <dnpd>
This command sets the specified phase's voltage channel flicker duty cycle value for
rectangular modulation.
The query command will return the present duty cycle value. The returned number will
be in standard scientific format (10.55 would be returned as 1.055E1).

5-61. Extended flicker sub-system

The extended flicker sub-system allows the generation of signals conforming to the
flicker-meter test scenarios described in IEC 60000-4 section 4.3.3 to 4.3.7.
When one of the extended functions is active, it will over-ride the existing behaviour of
the phenomena selected for the voltage channel of the passed phase (the current channel
will be unaffected), the original settings will be restored on disabling the extended
function.

5-62. Extended flicker state

[SOURce]:PHASe<x>:VOLTage:EFLicker:[STATe](?) <bool> { ON | OFF | 0 | 1 }
Enable or disable the currently selected extended flicker signal configuration (for the
passed phase x). If the output is on, the new configuration will be applied immediately to
the output.

5-63. Configure signal

[SOURce]:PHASe<x>:VOLTage:EFLicker:CONFigure(?) <cpd> { PF | MZ | HS |
PJ | RV }
Select the extended flicker signal function (for the passed phase x).
Note: The instrument's state will not change until ':EFLicker:STATe' is set to 'ON' (or
'1' ). If the extended flicker function is changed, ':EFLicker:STATe' will automatically
be set back to 'OFF' or ('0').
The extended functions are:
The default value is 'PF' (periodic frequency changes).
5-36
PF - Flicker signal with periodic frequency changes (section 4.3.3 of IEC 601000-4).
MZ - Distorted voltage with multiple zero crossings (section 4.3.43 of IEC 601000-
4).
HS - Harmonics with sideband (section 4.3.5 of IEC 601000-4).
PJ - Phase jumps (section 4.3.6 of IEC 601000-4).
RV - Rectangular voltage changes with duty cycle (section 4.3.7 of IEC 601000-4).