PROBE MAINTENANCE

PERIODIC MAINTENANCE
Inspect the probe and the cable. The cable used for the connection to
the meter must be intact and there must be no points of broken
insulation on the cable or cracks on the probe stem or bulb.
Connector must be perfectly clean and dry. If any scratches or cracks
are present, replace the electrode. Rinse off any salt deposits with
water.
CLEANING PROCEDURE
For better accuracy in measurements and to ensure a good perfor-
mance of the probe, a frequent cleaning is recommended.
For this purpose, soak it in Hanna HI 700661 Cleaning Solution for
5 minutes.
Note: For particular dirty (as for example protein, oil or grease) see
the "Accessories" section for Hanna specific solutions.
Note: After cleaning the probe, it is recommended to recalibrate the
meter. If it is not possible to calibrate, the probe has to be
replaced with a new one.
Note: For field applications, it is always recommended to keep a
spare probe handy. When anomalies are not resolved with
simple maintenance, change the probe and recalibrate the
meter.
12

ACCESSORIES

PROBES
HI 1285-0 Combination, amplified pH/EC/TDS probe with built-in
temperature sensor, 8-pin DIN connector and 1 m
(3.3') cable
HI 1285-5
Combination, amplified pH/EC/TDS/temperature probe
with built-in temperature sensor, 8-pin DIN connector
and 1 m (3.3') cable
pH BUFFER SOLUTIONS
HI 7004L
pH 4.01 buffer solution, 500 mL bottle
HI 7006L
pH 6.86 buffer solution, 500 mL bottle
HI 7007L
pH 7.01 buffer solution, 500 mL bottle
HI 7009L
pH 9.18 buffer solution, 500 mL bottle
HI 7010L
pH 10.01 buffer solution, 500 mL bottle
CONDUCTIVITY & TDS CALIBRATION SOLUTIONS
HI 7031L
1413 µS/cm solution, 500 mL bottle
HI 7032L
1382 ppm (mg/L) solution, 500 mL bottle
OTHER SOLUTIONS
HI 700661P Cleaning Solution, 20 mL sachet (25 pcs)
HI 70300L Storage Solution, 500 mL bottle
HI 7073L
Protein Cleaning Solution, 500 mL bottle
HI 7074L
Inorganic Cleaning Solution, 500 mL bottle
HI 7077L
Oil & Fat Cleaning Solution, 500 mL bottle
OTHER ACCESSORIES
ChecktempC Electronic thermometer (-50.0 to 150.0°C)
HI 710015 Shockproof rubber boot, blue
HI 710016 Shockproof rubber boot, orange
HI 710022 Blue protective case
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.