SPECIFICATIONS OF
HI 93510 & HI 93510N
SPECIFICATIONS
Range
Resolution
0.1°F (-58.0 to 230.0°F) / 0.2°F (outside)
Accuracy
for one year, excluding probe error
Typical EMC
Deviation
Battery
3x1.5V AA (IEC LR6) batteries,
approx. 2000 hours of continuous use (BL off)
Auto-off
user selectable: 8 min, 60 min, disabled
(HI 93510 only)
Environment
-10 to 50°C (14 to 122°F); RH 100%
Dimensions
150 x 80 x 36 mm (5.9 x 3.1 x 1.4")
Weight
Main features:
• HI 762 probes series
• Available HI 762 accuracy test plugs
• Display of temperature in degrees Celsius or Fahrenheit
• Range -50 to 150°C or -58 to 302°F
• HOLD function
• HI and LO (Max and Min) temperature values always displayed
on the LCD
• Remaining battery life indication / low battery detection
• Waterproof casing
• Auto-off capability (HI 93510 only)
• Backlight feature (HI 93510N only)
• User calibration at 0°C (HI 93510N only)
-50.0 to 150.0°C
-58.0 to 302.0°F
0.1°C
±0.4°C / ±0.8°F
±0.5°C / ±1°F
with HI 762BL probe
235 g (8.3 oz.)
4
HI 93510
Keyboard Functions:
ON/OFF : turn the meter ON and OFF.
HOLD : freeze the reading on display.
°C/°F : change reading unit (°C or °F).
CLR : clear the HI and LO values.
BL (HI 93510N only) : toggle the backlight ON and OFF.
CAL (HI 93510N only) : press and hold for 5 seconds to enter the
Calibration mode (with reading within ±3°C range); press to
abort the Calibration mode.
HI 93510N
5