Performing voltage measurements

Ensure that the terminal connections are connected correctly before making any
C A U T I O N
measurement. To avoid damaging the multimeter, do not exceed the rated input limit.
AC voltage
• Five ranges: 500.00 mV, 5.00 V, 50.00 V, 500.00 V, 750.00 V
• Measurement method: AC coupled true RMS—measure the AC component with up to 400 VDC
• Crest factor: Maximum 3:0 at full scale
• Input impedance: 1 MΩ±2%parallel with capacitance <100 pF on all ranges
• Input protection: 750 V RMS on all ranges
1 Press
2 Connect the red and black test leads to the respective input terminals
3 Probe the test points and read the display. In autoranging mode, the

Figure 1-10 ACV terminal connection and display

18
bias on any range
.
as shown in
Figure 1-
multimeter automatically selects the appropriate range and the
measurement is displayed.
+
ACV
voltage
source
10.