Fluke RPM4-AD Operation And Maintenance Manual page 101

Reference pressure monitor, air data version
Table of Contents

Advertisement

CONTn
Purpose
Have the RPM4-AD reply continuously as each new measurement becomes available.
Query
"CONTn?"
Classic
"CONTn"
Optional Suffix
"n"
The active Q-RPT is assumed if no suffix is given.
'1'
'2'
'3'
Remarks
This command is somewhat different than most as it enables the RPM4-AD to continuously output
the selected Q-RPT's measurement without further queries. This continues until the next query is
received by the RPM4, or if the ABORT" or "*CLS" command is received. The frequency of the
measurements is determined by the readrate of the Q-RPT. The data returned includes the
pressure of the specified or active Q-RPT, the Q-RPT's unit of measure and measurement mode.
Example
Query sent:
(enhanced)
Query reply:
Example
Query sent:
(classic)
Reply:
See Also
"PR", "READRATE"
DATE
Purpose
Read or set the RPM4-AD date.
Command
"DATE date"
Query
"DATE?"
Classic
"DATE= date"
"DATE"
Arguments
date:
Remarks
The RPM4-AD has an internal real time calendar clock. The span of acceptable dates ranges
from 19800101 to 20791231. The reply is always in the YYYYMMDD format.
Example
Cmd sent:
(enhanced)
Query reply:
Example
Cmd sent:
(enhanced)
Query reply:
Example
Sent:
(classic)
Reply:
Errors
ERR# 6:
See Also
"TIME"
3.5.5.3
ERR
Purpose
Read the new available error message from the Error Queue.
Query
"ERR?"
Classic
"ERR"
Remarks
This program message obtains additional details about an error that has occurred. If the user
receives an "ERR# nn" reply, or the enhanced mode is enabled using the IEEE-488 interface and
an error has been detected, the error is put into a FIFO Error Queue. The "ERR" program
message pulls and replies the oldest error message available. In "classic" mode only the most recent
error can reside in the queue. "OK" is replied if there are no error messages left. If an error occurs in
enhanced mode using the IEEE-488 interface, no reply is generated even if the command was a
query.
Example
Query sent:
(enhanced):
Query reply:
Example
Sent:
(classic)
Reply:
See Also
4.4.2
Specify the Hi Q-RPT if the HL RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
Specify the Lo Q-RPT if the HL Q-RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
"CONT1?"
"1936.72 kPa a"
(repeats every measurement cycle)
"CONT"
"1936.72 kPa a"
(repeats every measurement cycle)
The date in the numerical only format "YYYYMMDD"
"DATE 20030115"
"20030105"
(No reply if GPIB-488)
"DATE? 20030105"
"20030105"
"DATE=20021201"
"20021201"
Missing or improper program message argument(s).
"ERR?"
"Numeric argument missing or out of range"
"ERR"
"Numeric argument missing or out of range"
Page 91
4. REMOTE OPERATION