PG7607™ OPERATION AND MAINTENANCE MANUAL
1.2
Specifications
1.2.1
General Specifications
Specifications below are specific to PG7607. All other specifications are identical to PG7000
specifications found in the PG7000 Operation and Maintenance Manual (P/N 3152117).
Instrument platform with no
Overall Pressure Ranges
Maximum Mass Load
Pressure Connections
Piston Controlled Clearance
Vacuum Reference Ports
1.2.2
Piston-Cylinder Assembly
The piston is mounted in a fixed position on the mounting post. The cylinder floats and
is rotated.
1.2.3
Pressure Measurements
The piston is mounted in a fixed position on the mounting post. The cylinder floats and
is rotated.
1 Sensitivity: The smallest variation in input detectable in output.
2 Reproducibility: Combined long term stability of piston-cylinder effective area and masses.
© 2010 Fluke Calibration
17 kg (29 lb)
Weight
mass loaded.
Gauge: 5 to 175 kPa (0.7 to 25.4 psi)
with 35 kg mass set
Absolute: 5 to 175 kPa (0.7 to 25.4 psi)
Clean, dry, non-corrosive gas
Operating Medium
35 kg (38 kg with AMH automated mass handler)
DH200
TEST Port
DH200
Bell Jar Vent Port
DH200
Pressure (CCP)
NOTE: DH200 is a gland and collar type fittings for 1/4 in.
(6 mm) coned and left hand threaded tubes equivalent to
AE SF250C, HIP LF4, etc.
KF25 on platform and bell jar (KF40 available on optional
AMH automated mass handler)
CE Conformance
Available, must be specified.
Tungsten carbide
Piston Material:
Cylinder Material:
Tungsten carbide
50 mm
Nominal Diameter:
Nominal Area:
2 000 mm
Free deformation with controlled clearance pressure
Mounting System:
(CCP) available on inside of piston.
< 0.1 mm/min
Typical Drop Rate:
Full mass load
1
0.01 Pa + 0.5 ppm
Sensitivity
:
2
± 2 ppm
Reproducibility
:
2