EUTECH INSTRUMENTS ECOTESTR EC HIGH Operating Instructions

High waterproof pocket tester
Hide thumbs Also See for ECOTESTR EC HIGH:

Advertisement

Quick Links

Operating Instructions
Before First Use: Soak sensor for 2 minutes in alcohol to remove stains/oils. Rinse in
clean water.
Testing:
1. Remove cap and press
to turn on the tester
2. Dip sensor fully into the test solution
3. Stir once and let the reading stabilize. Note the measured value
4. To hold reading, press
. Screen fl ashes
with blinking unit (mS) to indicate that tester is in the hold mode. Press
HC
cancel hold mode (
)
5. Press
to shut the tester off
Note: Tester automatically shuts off after 8.5 minutes of non-use to conserve batteries.
Calibration: The tester is factory calibrated. However, to ensure accuracy, regular
calibration is recommended.
1. Soak sensor for 10 minutes in alcohol to remove stains/oils. Rinse in clean water and
blot sensor dry
2. Press
to turn on the tester
3. Dip sensor fully into standard solution and wait for the value to stabilize
4. Press
to enter the calibration sequence. Display fl ashes
then shows a fl ashing default reading
5. Press
to scroll to the standard solution value
6. Release
and wait for the auto confi rmation
7. To abort calibration, press
to escape (
8. You have successfully calibrated the instrument
Maintenance: Clean the stainless steel sensor by periodically rinsing them in alcohol
for 10 - 15 minutes. Rinse sensor in clean water after each use.
Useful Notes: To avoid cross contamination, rinse between samples and calibration
standards with de-ionized water.
EcoTestr EC High Waterproof Pocket Tester
HO
once, then displays measurement
again to
CAL
momentarily and
ESC
)
Battery
cover
LIFT
Step 1
Lift and hold
Changing Batteries:
1. To remove batteries, lift up the front battery cover to release the front catch. Push up
and hold in that position
2. Next, lift the pocket clip to release the back catch and slide upwards to remove the
battery cover (Figure 1)
3. Replace old batteries with fresh ones. Note polarity as shown in battery compartment
4. No re-calibration is required after battery change. The tester's non-volatile memory
function retains calibration settings
Error Messages:
bA
1.
– Batteries are weak and need replacement
Er
2.
– Calibration is performed in solution value below 5 % of full-scale range.
Use higher value calibration standard
– Faulty temperature sensor
Or
3.
– Over range signal for measurements performed above the full-scale range
Reset Option: Reset option allows you to restore the calibration back to factory
default settings.
1. Press
to turn off tester
2. Press and hold
, then press
to turn the tester on. The display shows a
rS
fl ashing
(reset)
En
3. Press
again to confi rm reset (
) or press
Warranty: This tester is supplied with a warranty against manufacturing defects for a
period of ONE year from date of purchase.
Accessories:
Eutech
Oakton
Description
Order Code
Order Code
ECCON2764BT
00653-20
2764 µS/cm KCl calibration solution
ECCON1288BT
00606-10
12.88 mS/cm KCl calibration solution
LIFT
LIFT
Step 2
Lift and push up
Figure 1:
Removing battery cover
to change batteries
ESC
to escape (
) reset option