EUTECH INSTRUMENTS CYBERSCAN CON 1500 CONDUCTIVITY METER Instruction Manual

EUTECH INSTRUMENTS CYBERSCAN CON 1500 CONDUCTIVITY METER Instruction Manual

Bench conductivity/ tds/ resistivity meter

Advertisement

Quick Links

Instruction Manual
CyberScan CON 1500
Bench Conductivity/ TDS/ Resistivity Meter
Technology Made Easy ...
68X292335
Rev 1 01/04

Advertisement

Table of Contents
loading

Summary of Contents for EUTECH INSTRUMENTS CYBERSCAN CON 1500 CONDUCTIVITY METER

  • Page 1 Instruction Manual CyberScan CON 1500 Bench Conductivity/ TDS/ Resistivity Meter Technology Made Easy ... 68X292335 Rev 1 01/04...
  • Page 3 CyberScan CON 1500 bench meter as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Eutech Instruments’ Authorised Distributors or call us at (65) 6778-6876 for Eutech Instruments’ Customer Service Dept. for assistance.
  • Page 4: Table Of Contents

    TABLE OF CONTENTS 1. INTRODUCTION 2. GETTING STARTED 2.1. Connectors 3. USING THE METER 3.1. Conductivity Probes 3.2. Display/ Keys 3.3. Screen Display 4. CONDUCTIVITY OPERATION 4.1. Using Setup 4.2. Overview of Setup Pages in CyberScan CON 1500 4.2.1 Setup Page 1.0: View the Cal data 4.2.2 Setup Page 2.0: Set cell constant 4.2.3...
  • Page 5: Introduction

    Instruction Manual CyberScan CON 1500 INTRODUCTION Thank you for selecting the Eutech Instruments CyberScan CON 1500 bench-top meter. This instruction manual describes the operation of the meter. The state-of-art meter that you have purchased is easy to operate and will guide you through the various functions by displaying easy to understand prompts.
  • Page 6: Getting Started

    Instruction Manual CyberScan CON 1500 GETTING STARTED 2.1. Connectors 1. Review the layout and arrangement of the rear connector panel.
  • Page 7 Instruction Manual CyberScan CON 1500 2. Connect the electrode arm to the base plate. 3. Connect the power adapter’s output power jack to the meter’s rear panel DC input power socket and plug in the adapter to a power source.
  • Page 8: Using The Meter

    3. Connect the ATC probe by plugging the 1/8” mini-phone jack into the ATC socket located at the rear panel of the meter. The Eutech Instruments 4-cell probe has a built-in temperature sensor and therefore a separate ATC probe is not required.
  • Page 9 Instruction Manual CyberScan CON 1500 1. Rinse the conductivity probe sensing elements with distilled or deionised water between samples. For long term storage, the probes can be stored dry or in distilled water. Note that the cell constant may change slightly due to storage or use, and it must be re-evaluated with the use of standard conductivity solution (standardisation) prior to use.
  • Page 10: Display/ Keys

    Instruction Manual CyberScan CON 1500 3.2. Display/ Keys Overview of the meter screen display and function key layout. Press std key to initiate standardisation. Or press std key at the ‘Standardize’ mode allows you to exit and return to measurement mode without accepting the calibration.
  • Page 11: Screen Display

    Instruction Manual CyberScan CON 1500 3.3. Screen Display Familiarise yourself with the layout of the digital screen display.
  • Page 12: Conductivity Operation

    Instruction Manual CyberScan CON 1500 CONDUCTIVITY OPERATION 4.1. Using Setup The setup button brings you to the setup mode of the various parameters. Press setup while in measurement to access to the setup mode. The ▲/save or ▼/view keys allow you to scroll through the setup pages available.
  • Page 13: Overview Of Setup Pages In Cyberscan Con 1500

    Instruction Manual CyberScan CON 1500 4.2. Overview of Setup Pages in CyberScan CON 1500 Each mode selection (conductivity, resistivity or total dissolved solids) has similar setup pages except for TDS mode which includes Setup P6.0 for setting TDS factor. Refer to Overview of Setup Pages shown below.
  • Page 14: Setup Page 1.0: View The Cal Data

    Instruction Manual CyberScan CON 1500 4.2.1 Setup Page 1.0: View the Cal data This setup page allows you to view the calibrated buffer value depending on the display mode (Conductivity/ Resistivity/ TDS) selected, the electrode type (2-Cell/ 4-Cell), respective calibrated electrode cell constant/s and the calibrated buffer value.
  • Page 15 Instruction Manual CyberScan CON 1500 3. Press enter repeatedly to view all previous calibration data, starting from the first range till fourth range. The meter will display the calibration value of that particular range provided calibration is performed. If any of the range is not being calibrated the upper display shows “---“.
  • Page 16 Instruction Manual CyberScan CON 1500 4. When you have scrolled through all the cell constants, you will automatically come to the screen shown below. This screen displays the meter’s calibrated value. 5. Press enter key to go to the next setup page P2.0. Press std key if you wish to return to measurement mode.
  • Page 17: Setup Page 2.0: Set Cell Constant

    Instruction Manual CyberScan CON 1500 4.2.2 Setup Page 2.0: Set cell constant This setup page allows you to set the cell constant value as 0.1, 1.0 or 10.0. 1. Access the Set cell constant setup page by pressing setup during any measurement mode screen. 2.
  • Page 18 Instruction Manual CyberScan CON 1500 3. The cell constant can be selected as 0.1, 1.0 and 10.0 using the ▲/save or ▼/view keys to scroll through. 4. Press enter key to confirm selection OR press std to exit from this setup page. 5.
  • Page 19: Setup Page 3.0: Set Temperature Coefficient

    Instruction Manual CyberScan CON 1500 4.2.3 Setup Page 3.0: Set Temperature Coefficient This setup page allows you to set the temperature coefficient in the range of 0.000 to 10.000%. 1. Access the Set Temperature Coefficient setup page by pressing setup during any measurement mode screen. 2.
  • Page 20 Instruction Manual CyberScan CON 1500 3. Press enter to view the current temperature coefficient value. 4. The temperature coefficient value can be adjusted in the range of 0.000 to 10.000% using the ▲/save or ▼/view keys. COEFF 5. Press enter key to confirm selection OR press std to exit from this page.
  • Page 21: Setup Page 4.0: Set The Normalisation Temperature

    Instruction Manual CyberScan CON 1500 4.2.4 Setup Page 4.0: Set the Normalisation Temperature This setup page allows you to set the Normalisation temperature value in the range of 15.0 to 30.0°C. 1. Access the Set the Normalisation Temperature setup page by pressing setup during any measurement mode screen.
  • Page 22 Instruction Manual CyberScan CON 1500 3. Press enter key to view the current normalisation temperature value. 4. Press the ▲/save or ▼/view keys to adjust the normalisation temperature value in the range of 15.0 to 30.0°C. °C 5. Press enter key to confirm the value set OR press std to exit from this page.
  • Page 23: Setup Page 5.0: Set The Auto Calibration Mode (Acal For Conductivity Mode Only)

    Instruction Manual CyberScan CON 1500 4.2.5 Setup Page 5.0: Set the Auto Calibration mode (ACAL for Conductivity mode only) This setup page allows you to select an auto calibration or a manual calibration. This option is only applicable during conductivity mode of measurement.
  • Page 24 Instruction Manual CyberScan CON 1500 3. Press the enter key to go into the Set the Auto Calibration Mode (ACAL) setup. 4. Use the ▲/save or ▼/view keys to select the options of YES (Auto) or NO (Manual) so as to configure the Calibration Mode to Auto Calibration or Manual Calibration mode.
  • Page 25: Setup Page 5.0: Set The Calibration Mode (Spcal For All Modes)

    Instruction Manual CyberScan CON 1500 4.2.6 Setup Page 5.0: Single/Multi Point Calibration mode (SPCAL for all modes) This setup page allows you to select a single-point or a multi-point calibration. This option is applicable during all modes of measurement (Conductivity, Resistivity and Total Dissolved Solids). 1.
  • Page 26 Instruction Manual CyberScan CON 1500 3. Press the enter key to go into the Set the Single/Multi Point Calibration Mode (SPCAL) setup page. 4. Use the ▲/save or ▼/view keys to select the options of YES (single-point) or NO (multi-point) so as to configure the Calibration Mode to Single-point Calibration or Multi-point Calibration mode.
  • Page 27: Setup Page 6.0: Set The Tds Factor. (Only Applicable In Tds Mode)

    Instruction Manual CyberScan CON 1500 4.2.7 Setup Page 6.0: Set the TDS Factor. (Only applicable in TDS mode) This setup page allows you to set the TDS Factor in the range of 0.40 to 1.00. This option is only applicable during Total Dissolved Solids mode of measurement.
  • Page 28 Instruction Manual CyberScan CON 1500 3. Press the enter key to go into the Set the TDS Factor setup and view the current TDS Factor. 4. The TDS Factor of 0.40 to 1.00 can be set using the ▲/save or ▼/view keys.
  • Page 29: Setup Page 7.0: Set The Temperature Unit

    Instruction Manual CyberScan CON 1500 4.2.8 Setup Page 7.0: Set the Temperature Unit. This setup page allows you to select unit of measure for Temperature either in °C or °F. To Select Temperature Unit Access the Set the Temperature Unit menu during any measurement mode screen by pressing the setup key.
  • Page 30 Instruction Manual CyberScan CON 1500 Press the enter key to go into the Set the Temperature Unit setup page and view the current temperature unit. Use the ▲/save or ▼/view key to choose either °C or °F. select °C view save select °F...
  • Page 31: Setup Page 8.0: Set The Stability Indicator

    Instruction Manual CyberScan CON 1500 4.2.9 Setup Page 8.0: Set the Stability Indicator This setup page allows you to activate/ de-activate the stability indicator. Once activated, the STABLE icon appears on the display when reading stabilises. Access the Set the Stability Indicator setup page during any measurement mode screen by pressing the setup key.
  • Page 32 Instruction Manual CyberScan CON 1500 Press the enter key to go into the Set the Stability Indicator setup. Use the ▲/save or ▼/view keys to toggle between the options of YES (enable stability indicator) and NO (disable stability indicator). view save 5.
  • Page 33: Setup Page 9.0: Set The Baud Rate

    Instruction Manual CyberScan CON 1500 4.2.10 Setup Page 9.0: Set the Baud Rate This setup page allows you to set the baud rate (bits per second) of the communication protocol interface. Access the Set the Baud Rate setup page during any measurement mode screen by pressing the setup key.
  • Page 34 Instruction Manual CyberScan CON 1500 Press the enter key to go into the Set the baud rate page and view the current baud rate value. The baud rate can be set as 4800, 9600, 19200 or 38400 bps using the ▲/save or ▼/view keys. Print P rint press...
  • Page 35: Setup Page 10.0: Set The Parity Bit

    Instruction Manual CyberScan CON 1500 4.2.11 Setup Page 10.0: Set the Parity Bit This setup page allows you to set the parity bit of the communication protocol interface. 1. Access the Set Parity Bit setup page from any measurement mode screen by pressing the setup key. 2.
  • Page 36 Instruction Manual CyberScan CON 1500 3. Press the enter key to go into the Set Parity Bit setup page. 4. Use the ▲/save or ▼/view keys to toggle between 0 (none), 1 (odd) or 2 (even). view save view save 5.
  • Page 37: Setup Page 11.0: Set The Stop Bit

    Instruction Manual CyberScan CON 1500 4.2.12 Setup Page 11.0: Set the Stop Bit This setup page allows you to set the stop bit of the communication protocol interface. To Set Stop Bit 1. Access the Set Stop Bit setup page from any measurement mode screen by pressing the setup key.
  • Page 38 Instruction Manual CyberScan CON 1500 3. Press the enter key to go into the Set Stop Bit setup page. 4. Use the ▲/save or ▼/view keys to toggle between 1 or 2. view save 5. Press enter key to confirm selection OR press std to exit from this page.
  • Page 39: Setup Page 12.0: Select The Print Option

    Instruction Manual CyberScan CON 1500 4.2.13 Setup Page 12.0: Select the Print option This setup page allows you to print current displayed data or data stored in the meter’s memory to a computer or printer via its RS232 interface port. Note all the communication protocol for both the meter and computer/printer must match before successful printing can be performed.
  • Page 40 Instruction Manual CyberScan CON 1500 3. Press the enter key to go into the Select Print Data setup page. 4. Use the ▲/save or ▼/view keys to toggle between Current or Memory data print out selection. Print P rint view save Print P rint...
  • Page 41: Setup Page 13.0: Clear Stored Data

    Instruction Manual CyberScan CON 1500 4.2.14 Setup Page 13.0: Clear Stored Data This setup page allows you to clear all stored data sets (from previous measurements) in the meter’s memory for new data to be stored. Note old data sets will be overwritten by any new data sets in the event when the stored locations have exceeded.
  • Page 42 Instruction Manual CyberScan CON 1500 3. Press the enter key and use the ▲/save or ▼/view keys to toggle between NO or YES. Select YES to clear all stored data sets or NO if no change is to be made. view save 4.
  • Page 43: Setup Page 14.0: Clear User Calibration

    Instruction Manual CyberScan CON 1500 4.2.15 Setup Page 14.0: Clear User calibration This setup page allows you to clear the user calibrated values. This option clears respective user calibration depends on the mode you are in. (e.g.: If you access the setup menu from conductivity measurement mode, then this option only clears the conductivity user calibration) 1.
  • Page 44 Instruction Manual CyberScan CON 1500 3. Press the enter key to make the selection and use the ▲/save or ▼/view keys to toggle between NO or YES. Select YES to clear all the existing buffer values or NO if no change is to be made. view save 4.
  • Page 45: Setup Page 15.0: Replatinisation Process

    Instruction Manual CyberScan CON 1500 4.2.16 Setup Page 15.0: Replatinisation Process This setup page allows you to replatinise the probe and is only applicable for 2-cell probes. Replatinisation is the process of replacing the platinum on the surfaces of the 2-cell conductivity probes that may flake or wear off over time.
  • Page 46 Instruction Manual CyberScan CON 1500 1. Immerse the probe into a suitable replatinising solution. 2. Access the Replatinisation Process setup page from any measurement mode screen by pressing the setup key. 3. Use the ▲/save or ▼/view keys to scroll until the screen displays the as shown below.
  • Page 47 Instruction Manual CyberScan CON 1500 4. Use the ▲/save or ▼/view keys to scroll through the options of YES and NO for the selection of replatinisation. 5. Press the enter key to confirm selection OR press std to exit from this page. view save 6.
  • Page 48: Standardisation

    Instruction Manual CyberScan CON 1500 STANDARDISATION Conductivity probes are generally identified as having a characteristic cell constant, 0.1, 1.0 or 10.0 which reflects their physical geometry and their range of application. However, these are typically nominal cell constants. The actual cell may vary somewhat from the nominal values, and therefore the actual cell constant must be calculated using a solution with a known conductivity value.
  • Page 49: Conductivity Standardisation (Manual)

    Instruction Manual CyberScan CON 1500 5.1. Conductivity standardisation (Manual) 1. Immerse the electrode into the sample solution. Stir moderately. Make sure that the meter is in the measurement mode. If you are using the 2-cell electrode, and separate temperature probe is NOT available, the meter will take 25°C as the default temperature.
  • Page 50 Instruction Manual CyberScan CON 1500 5. After adjusting to the std solution value (which will be displayed at the upper display), press enter to initiate the standardisation and on successful completion, meter returns to measurement mode Press std to exit from the standardisation page. Adjustable standardisation window provided is ±40% of current reading.
  • Page 51: Conductivity Standardisation (Auto)

    Instruction Manual CyberScan CON 1500 5.2. Conductivity standardisation (Auto) Refer to Section 4.2.5: Set the Auto Calibration Mode (ACAL) on Page 19 to enable the automatic calibration of conductivity feature in this meter. Standard buffer values: 84.0 uS/ 1.413 mS/ 12.88 mS/ 111.8 mS 1.
  • Page 52 Instruction Manual CyberScan CON 1500 STABLE press enter to accept STD Solution...
  • Page 53: Resistivity Standardisation (Manual)

    Instruction Manual CyberScan CON 1500 5.3. Resistivity Standardisation (Manual) 1. Immerse the electrode into the sample solution. Stir moderately. Make sure that the meter is in the measurement mode. If you are using the 2-cell electrode, and separate temperature probe is NOT available, the meter will take 25°C as the default temperature.
  • Page 54 Instruction Manual CyberScan CON 1500 5. After adjusting to the std solution value (which will be displayed at the upper display), press enter to initiate the standardisation and on successful completion, meter returns to measurement mode Press std to exit from the standardisation page. Adjustable standardisation window provided is ±40% of current reading.
  • Page 55: Tds Standardisation (Manual)

    Instruction Manual CyberScan CON 1500 5.4. TDS Standardisation (Manual) 1. Immerse the electrode into the sample solution. Stir moderately. Make sure that the meter is in the measurement mode. If you are using the 2-cell electrode, and separate temperature probe is NOT available, the meter will take 25°C as the default temperature.
  • Page 56 Instruction Manual CyberScan CON 1500 5. After adjusting to the std solution value (which will be displayed at the upper display), press enter to initiate the standardisation and on successful completion, meter returns to measurement mode Press std to exit from the standardisation page. Adjustable standardisation window provided is ±40% of current reading.
  • Page 57: Temperature Calibration

    Instruction Manual CyberScan CON 1500 TEMPERATURE CALIBRATION If the temperature probe is not connected, the meter will display 25.0°C default icon on the secondary display. If the temperature probe is connected, then meter will sense the actual temperature and display the ATC icon. From any measurement mode (Conductivity/ Resistivity / Total Dissolved Solids):- 1.
  • Page 58 Instruction Manual CyberScan CON 1500 Depending on which temperature probe is connected (4-cell/ 2-cell temperature probe), user can adjust the temperature value using the ▲/save or ▼/view keys. [Window: ± 5°C (9°F)] If no probe is connected, user can adjust the default temperature value.
  • Page 59: Measurement

    Instruction Manual CyberScan CON 1500 MEASUREMENT 7.1. Conductivity Measurement 1. Immerse the electrode into the sample solution. Stir moderately. 2. When the meter senses that the reading has stabilised, the stable icon will appear under the reading. The reading may be recorded at this time.
  • Page 60: Resistivity Measurement

    Instruction Manual CyberScan CON 1500 7.2. Resistivity Measurement Unit will be in ohm, kohm or Mohm. 1. Immerse the electrode into the sample solution. Stir moderately. 2. When the meter senses that the reading has stabilised, the stable icon will appear under the reading. The reading may be recorded at this time.
  • Page 61: Tds Measurement

    Instruction Manual CyberScan CON 1500 7.3. TDS Measurement TDS readings will be in ppm unit. TDS factor can be selected from the setup menu. (0.40 to 1.0). Default: 0.66. 1. Immerse the electrode into the sample solution. Stir moderately. 2. When the meter senses that the reading has stabilised, the stable icon will appear under the reading.
  • Page 62: Memory

    Instruction Manual CyberScan CON 1500 8. MEMORY 8.1. Store Value into Memory In any measurement mode, press ▲/save key to store the displayed save reading into the meter’s non-volatile memory. A memory location is shown momentarily and the meter returns to measurement mode. 8.2.
  • Page 63: Print Data

    Instruction Manual CyberScan CON 1500 9. PRINT DATA 9.1. Printing Data Depending on the print option in the meter setup, pressing the print key allows you to print either current displayed reading or stored data from meter to a PC or printing device via a RS232 communication print cable.
  • Page 64: Conductivity Theory

    Instruction Manual CyberScan CON 1500 10. CONDUCTIVITY THEORY Conductance is a metric associated with the ability of primarily aqueous solutions to carry an electrical current, I, between two metallic electrodes when a voltage E is connected to them. Though water itself is a rather poor conductor of electricity, the presence of ions in the water increases its conductance considerably, the current being carried by the migration of the dissolved ions.
  • Page 65 Instruction Manual CyberScan CON 1500 backwards- however, this term has been replaced by the term “siemen”. Conductivity measurements are reported as Siemens/ cm, since the value is measured between opposite faces of a cell of a known cubic configuration. With most aqueous solutions, conductivity quantities are most frequently measured in microSiemens per cm (µS/ cm) or milli-Siemens per cm (mS/ cm).
  • Page 66 Instruction Manual CyberScan CON 1500 Conductivity Measurement CyberScan conductivity probes consist of glass or epoxy bodies in which platinum or platinised sensing elements and are designated two-cell electrode has two such sensing elements and are designated two-cell electrodes. The previous discussion has focused on this type of electrode.
  • Page 67 25°C. The meter requires no regular maintenance, but it is recommended to occasionally wipe down the front with a damp cloth. If there are any further questions regarding maintenance, contact Eutech Instruments at (65) 6778 6876.
  • Page 68: Cleaning

    Instruction Manual CyberScan CON 1500 11. CLEANING This meter requires no regular maintenance, but it is recommended to occasionally wipe down the front with a damp cloth from time to time. 12. TROUBLESHOOTING The CyberScan CON 1500 displays pertinent error messages to guide you should an error occur with a measurement or meter operation.
  • Page 69: Warranty

    Instruction Manual CyberScan CON 1500 13. WARRANTY Eutech Instruments supplies this bench meter with a 3-year warranty and 6-month warranty for electrode against manufacturing defects from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warranty period, please return, freight prepaid, and correction will be made without charge.
  • Page 70: Notice Of Compliance

    Warning! To meet or exceed FCC regulations and comply with CE requirements, the Eutech Instruments supplied power supply must be used. Use of a power supply that is not approved by Eutech Instruments may cause safety hazards and/or cause unit to exceed EMC limits and/or damage unit.
  • Page 71: Meter Specifications

    Instruction Manual CyberScan CON 1500 15. METER SPECIFICATIONS Conductivity Resistivity 0.000 to 500.0 mS/cm 0.000 to 99999 ppm 0.000 to 99.99 MΩ Range 0.000 to 200.0 uS 0.000 to 20.00KΩ 0.000 to 200.0 ppm Internal Range 200.0 uS to 2.000 mS 20.00K to 200.0K 200.0 ppm to 2000 ppm 2.000 mS to 20.00 mS...
  • Page 72: Accessories

    Instruction Manual CyberScan CON 1500 18 ACCESSORIES Consult your Authorised Distributors for these items and other range of specialised electrodes. Replacement Meter & Meter Accessories EC-CON1500/13S CyberScan CON 1500 Bench Conductivity/ Resistivity/ TDS with 4-cell epoxy bodied conductivity electrode, K=1.0 (EC- 620-165), integral electrode holder, Data Acquisition Software &...
  • Page 73 3’000 ppm 442 Calibration Solution (20 X 20ml) Note: COA and MSDS information can be downloaded from our homepage site at www.eutechinst.com under Tech-tips. For a complete selection of electrodes and accessories, please contact your Eutech Instruments nearest distributor or sales representative for details.
  • Page 74 NOTES...
  • Page 76 For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit our website listed below: Manufactured by: Distributed by: Eutech Instruments Pte Ltd. Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail: marketing@eutechinst.com...