CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide pH 1100 and pH 2100 Bench pH and Bench pH/Ion Meter 68X090822 Technology Made Easy ... Rev 2 02/04...
Page 3
CyberComm Pro version via our website at www.eutechinst.com/cybercomm- software.htm Eutech Instruments/ Oakton Instruments reserves the rights to change, make improvement and modify specifications without prior notice and cannot accept any responsibility for damage or malfunction to the instrument caused by improper use.
TABLE OF CONTENTS 1 DATA ACQUISITION IN WINDOWS VERSION The CyberComm DAS Package System Requirements 1.2.1 Setting the Communication Parameters 2 INSTALLATION OF CYBERCOMM DATA ACQUISITION SOFTWARE (DAS) Loading of CD-ROM User Name and Organisation Entry Splash Screen 3 CYBERCOMM PRO MAIN MENU Communication Settings (File) Data Log Options (File) Connect to Meter (File)
Instruction Manual CyberComm pH 1100/ 2100 INSTALLATION OF CYBERCOMM DATA ACQUISITION SOFTWARE (DAS) Loading of CD-ROM See Figure 1. Before running the program, check that the pH 1100/ pH 2100 meter is properly connected to the computer. It is preferable to connect to either one of the COM (serial) ports available on a computer. Note the number of the COM port that you have connected to the meter as this will be required in setting up the CyberComm DAS later.
Instruction Manual CyberComm pH 1100/ 2100 User Name and Organisation Entry The CyberComm program (as below) prompts you to enter the name of the user and organization. Figure 2: CyberComm Pro Installation/ User Information See Figure 2. Follow the instruction in the subsequent menu for your desired location of this file to be saved. Make relevant changes until the setup is completed (shown below).
Instruction Manual CyberComm pH 1100/ 2100 Splash Screen See Figure 4. Upon clicking the Finish button, a splash screen pops up (shown below). The meter’s communication protocol setting i.e. baud rate, parity and stop bits must be properly set before you power the meter on. DO NOT press any key on the meter while the program is running.
Instruction Manual CyberComm pH 1100/ 2100 CYBERCOMM PRO MAIN MENU Figure 5: CyberComm Main Menu See Figure 5. Select Communications Settings and set the appropriate requirements before connecting the meter. Communication Settings (File) See Figure 6. To make changes, use the drop-down menu of each parameter to set to desired values to match the meter’s settings.
Instruction Manual CyberComm pH 1100/ 2100 Data Log Options (File) See Figure 7. a) Time between successive readings (1-3600 sec). Default is 10 sec. Select the time lag preferred between each successful reading using the arrow keys. b) Length of Time to Capture Data. Default is 0 min. (0= indefinitely, units = min) Select the total length of time preferred to capture data using the arrow keys.
Instruction Manual CyberComm pH 1100/ 2100 MAIN WINDOW (AFTER CONNECTING METER TO COMPUTER) CyberComm Pro Manuals (Help) See Figure 8 and 9. The main menu appears with the File and Help drop down menus, each having a list function settings shown below.
Instruction Manual CyberComm pH 1100/ 2100 4.1.1 File Options (File) See Figure 10. Select File Options from the File drop down menu. Enter the relevant information for useful future references and type a file name/or file location that you wish to store these details. Click OK upon completion to save information.
Instruction Manual CyberComm pH 1100/ 2100 Sub menus & Function Icons on the CyberComm Program See Figure 12. The figure shows the CyberComm Pro software Main Menu after linking with the meter. From this menu, most of the meter key functions can be emulated except calibration of pH probe and CAL EDIT key function. This Sub Menu has following features: Figure 12: Running CyberComm Pro/ Complete Main Menu (Connected to meter)
Instruction Manual CyberComm pH 1100/ 2100 Function Icons See Figure 13. See also Figure 12. The function icons on the main menu each serve a vital role in the CyberComm Pro software, providing user friendly one click access. Figure 13: Function Icon Panel (extracted from Figure 11) Icon Function Name Description...
Page 16
Instruction Manual CyberComm pH 1100/ 2100 Stability definitions High stability: reading is stabilized quickly, but reading is less accurate. Medium stability: reading stability is averaged between high and low stability. Low stability: reading is stabilized slowly, but guarantees high accuracy Icon 11.
Page 17
Instruction Manual CyberComm pH 1100/ 2100 Icon 12. mV Settings See Figure 17. a) mV Alarm Settings Using the Low and High options, select the ‘safe’ range of readings. Any readings not within the selected range will trigger the alarm alert. b) mV Calibration Due Check to enable the calibration due warning or uncheck to disable Calibration due period can be set from 1 to 365 days to remind you...
Page 18
Instruction Manual CyberComm pH 1100/ 2100 Icon 13. Ion Settings (Only applicable for pH 2100) See Figure 18. a) Ion Settings Select Ion Mode from the drop down menu; Ppm, ppt, mg/l, g/l, mmol/l or mol/l b) Ion Resolution Select High or Low ion resolution mode. Select High when electrode output is ±500 mV value, Low when electrode output is ±1850 mV value.
Instruction Manual CyberComm pH 1100/ 2100 Icon 14. Meter Settings See Figure 19. Program your meter functions via this icon on CyberComm Program. a) Date and Time Click on the textbox then a calendar will appear for you to set the date. b) Backlighting Enable or disable the backlight and choose the On period from 0 to 10 minutes.
Instruction Manual CyberComm pH 1100/ 2100 4.3.5 Temperature Reading This curve shows the temperature in which the measurement is carried out simultaneously. You can choose to view the reading with or without the temperature curve by clicking on the ‘Show/Hide Temperature’ icon on the icon panel.
Instruction Manual CyberComm pH 1100/ 2100 PRINT READINGS Printing using a serial printer Connect the pH 1100/ 2100 meter to the serial printer via RS232C cable (refer to the list of accessories). The configuration of the RS232C port is as shown below. Pin Number Description Transmit Data...
Page 28
For more information on Eutech Instruments/ Oakton Instruments products, contact your nearest distributor or visit our website listed below: Eutech Instruments Pte Ltd. Oakton Instruments Distributed by: P.O Box 5136, Blk 55, Ayer Rajah Crescent, Vernon Hills, IL60061, USA #04-16/24 Singapore 139949...