Advanced Operation; Change Measurement Options - Hach PHC30101 User Manual

Refillable ph probe
Table of Contents

Advertisement

Advanced operation

Setting
Measurement Options
Calibration Options
Check Standard Options
Units

Change measurement options

6
8. If the display shows Check Standard Failed, the measurement is outside of accepted
limits set by the administrative user and a recalibration is recommended. If the
acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the probe
calibration and status indicator, calibrate the probe (refer to
Parameter-specific settings can be changed through the Full Access Options menu. Details
about menu navigation, available options and how to change them are given in the screens,
tables and procedures throughout this section.
The settings that can be changed are shown in
Table 1 Parameter-specific settings
Options
Resolution
Upper and lower range limits
Buffer set
Buffer set values (if Custom Buffer Set selected)
Reminder
Minimum Cal points
Slope limit
Standard (temperature compensated buffer or custom at 25 °C)
Reminder
Acceptance criteria
Standard value at 25 °C (if custom standard selected)
pH
mV
Methods are groups of default or user-defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt for a new name is shown after the changes are entered. The settings are saved with
this name to distinguish them from the default method settings, which cannot be changed.
A saved method can be used instead of multiple adjustments to the individual settings.
Changes made to a user-defined method are automatically saved with the existing name.
Multiple methods can be saved for the same probe on each meter.
1. Make sure a probe is connected to the meter.
2. Push
and select PHC301 Settings.
3. Select Modify Current Settings.
4. Select Units. Select pH (default) or mV.
Note: The mV option can be used to find the probe offset in a pH 7 buffer or to measure the slope.
Both units are shown when the detail display mode is selected.
Calibration
Table
1.
on page 3).
DOC022.53.80032

Advertisement

Table of Contents
loading

This manual is also suitable for:

Phc30103

Table of Contents