Table 1-1.
Specifications and Suppiementa! Characteristics (continued)
Typical output voltage overshoot:
Final value
Overshoot
<3 V
3 V
>3V
0.1V
Readback {25 ±5 °C)
Voltage
Resolution: 15 mV
Accuracy: 0.07% 4-50 mV
TC: 30 ppm 4-1 mV/°C
Current
Resolution: 2.5 mA
Accuracy: 0.20% 4-11 mA
TC: 70 ppm 4-0.5 mA /°C
OVP
Resolution: 37.5 mV
Accuracy: 0.25% 4-300 mV
TC: 100 ppm 4-5 mV
Front Panel Meters (25 ±5 °C)
VOLTS
Range
Resofutim
Accuracy
2 V
15 mV
0.07%
4
50 mV
20 V
15 mV
0.07% 4 55 mV
200 V
100 mV
0.07% 4100 mV
TC: Same as voltage readback
mp$
Range
Resolutkm
Accuracy
20 A
10 mA
0.20% 411 mA
TC: Same as current readback
OVP
Range: 2(X) V
Resolution: 100 mV
Accuracy: 0.3% 4-300 mV
TC: Same as OVP readback
Multiple Oprations
Up to two HP 6038A units may be connected in series or
auto-parallel, to provide increased output capabilities. Note
mixing supplies with dis-similar output capabilities is not
recommended; because under certain conditions, the lower
output unit may be stressed beyond it's maximum voltage
and/or current capabilities by the higher output unit.
Overvoltage Protection
Trip voltage adjustable via front panel control.
Range: 0-63 V
Resolution: 100 mV
Accuracy: 0.25% 4-300 mV (Set using DISPLAY
OVP
function}
TC: 100 ppm +5 mV/°C
Minimum setting above output voltage to avoid nuisance
tripping is 1.5 V.
Remote Analog Programming (25 ±5 ''O
Resistance programming:
0 to 4 k provides zero to maximum rated output voltage or
current. The resulting scale factor is 1 ohm/LSBfora 12-bit
system, which corresponds to 2(X) ohms/volt and 133
ohms/amp.
Voltage:
Accuracy: 0.42% 4-30 mV
TC; 50 ppm 4 3 mV
Current:
Accuracy: 0.8% 420 mA
TC: 100 ppm 4 0.3 mA
Voltage programming:
0 to 5 V provides zero to maximum rated output voltage or
current.
Voltage:
Accuracy: 0.25% 4 30 mV
TC: 10 ppm 43 mV
Current:
Accuracy: 0.15% 420 mA
TC: 70 ppm 40.3 mA
The programming inputs are protected against input voltages
up to ±40 V.
FAULT (FLT) )terminals 1 and 2)
(iol) low level output current = 4 1.25 mA max.
(Vo!) low level output voltage = 0.5 V max.
INHIBIT (INH) (terminals 3 and 4).
(Vih) high level input voltage = 2 V min.
(Vi!) low level input voltage = 0.5 V max.
(tw) Ptuse width = 50 ms min.
(td) Delay time = 50 ms max. (2.0 ms typ.)
(lil) low level input current = 1.25 mA max.
ABSOLUTE MAXIMUM RATINGS:
between FLT terminals (terminals 1 and 2): 4 17.5
Vdc max.
between INH terminal (terminals 3 and 4): 45.5
Vdc max.
between terminals 1 or 2 to pin 4 or chassis gound:
±42 Vdc.
Voltage and Current Monitor Outputs (25 ±5 °C}
0 to 5 V output indicates zero to maximum rated output.
All accuracy specifications are referred to the output of the
supply.
Voltage:
Accuracy: 0.25% 410 mV
TC; 10 ppm 42 mV /°C
Output impedance: 10.2 k fiohm ±5%
Current:
Accuracy: 0.15% 46 mA
TC: 70 ppm 4 0.3 mA/®C
Output impedance: 10.2 k 0 ±5%